USER'S GUIDE____________________________________________________________________
40 __________________________________________________________________ M211371EN-B
Output a Reading Once
Use the
SEND
command to output a single measurement message.
The output message is sent in the format that has been set using the
FORM
command.
SEND
[
address
]<cr>
where
address = Address of the transmitter, range 0 ... 255.
Specifying the address is only necessary if the target
transmitter is in POLL mode, and has not been accessed
with the OPEN command.
Example:
send
Tdf= 15.4 'C Tdfatm= 16.2 'C H2O= 18539 ppm P= 0.959
bara Pnorm= 0.951 bara Rhoo= 1.1 kg/m3 T= 25.4 'C
Configuring Measurement Parameters
Set Mixing Ratio
Use the
MIXRATIO
command to set the mixing ratio of SF
6
and the
other gas.
MIXRATIO
[
ratio
]<cr>
where
ratio = Amount of SF
6
in the measurement environment, range
0 ... 100%.
Example:
mixratio 50
MIX RATIO : 50.00 %