12
4. Allow at least 50 cm clearance to the sensor housing, to be able to take out the measuring sensor
for service.
5. The sensor needs a stable environment to act as a counterweight for the pendulum movement.
The support must prevent movement and excessive vibrations of the
ViscoSense
®
3
housing,
especially movement in the direction of the pendulum movement (see drawing- angular movement
around the sensor axis). Therefore there must be two supports to the inlet and outlet perpendicular
to the pipe and perpendicular to the
ViscoSense
®
3
. If one support would be used the lateral
vibrations would be converted to angular vibrations and the situation would be made worse instead
of better.
Figure 6 Sensor installation diagram
The angular position of the supports must always be 90° in relation to the sensor.
The distance from the flange to the support should be two times the diameter of the pipe.
ViscoSense
®
3 housing size
T. Max.
L. Max.
50 mm (2”)
350 mm
150 mm
65 mm (2 ½”)
400 mm
160 mm
80 mm (3”)
400 mm
165 mm
100 mm (4”)
450 mm
185 mm
Figure 7 Sensor installation diagram