Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999039 / 16/12/2009

Description VAP 5

The vacuum gauge VAP 5 relies on the fact that the thermal conductivity of the residual gas in the vacuum
chamber is a measure for the gas pressure. The thermal conductivity of gases is proportional to the
pressure within a certain range and is related to the molecular mass.
The thermal conductivity of  gases and vapours varies with their molecular mass. The gauge has been
adjusted for air (nitrogen) at the factory.

Pressure of gases with similar mass, such as O

2

 or CO, can be read off directly within the uncertainty

of the measurement.

With gases of lower or higher mass (H

2

, He, Ar, CO

2

) it is advisable to readjust the gauge head, primarily

at atmospheric pressure, using the gas to be measured, see section “Readjusting the gauge head“.

Description DVR 5

The DVR 5 is equipped with a ceramic pressure transducer to measure the actual pressure according to the
capacitive principle of measurement independent of the gas type and depending on the vacuum, i. e.
absolute. Electrically, the pressure transducer corresponds to a plate capacitor. A change in capacity
indicates a pressure change.
The pressure value is displayed with a resolution of 0.1 mbar in the range of 0.1 to 10 mbar  and with a
resolution of 1 mbar for pressures > 10 mbar.

Preparing for operation

Connect gauge head with connecting cable to the female connector at the rear side of  the equipment.
Screw the plug into the female connector.

Connect the gauge head to the vacuum chamber via the small flange connection or a hose nozzle.
Avoid contamination (oil/oil mist) of the gauge head when generating the vacuum with an oil-filled
vacuum pump.

Do not mount the gauge head directly at the pump. The diameter of the pipelines should be as large as
possible.

Switch on the equipment.

After switching on the vacuum gauge, an internal check and adjustment routine starts automatically
(indicated by the clock symbol and “CAL“ in the display).

How to change the pressure unit

The vacuum gauge comes switched to the mbar-scale (230 V version) or Torr-scale (110 V version). To
convert reading to mbar, Torr or hPa:

Press key-button by using a small screwdriver or the tip of a pencil (and hold) during switching on.

“Set“ and the pressure units (mbar, Torr, hPa) are displayed, the pressure unit as from last operation is
flashing.

Press key-button to change pressure unit. Approximately 2 sec. after the last tip the mode is termi-
nated.

Serial interface

The vacuum gauge 

DVR 5

 is equipped with a serial interface at the rear side of the housing.

Respectively plug-into or remove the cable from the interface (cable RS 232C, 9-pole, Sub-D) only if the
equipment is switched off.

The interface is not electrically isolated from the measuring circuit.

Содержание VAP 5

Страница 1: ...buted completely and unchanged It is strictly the users responsibility to check carefully the validity of this document with respect to his product manual no 999039 16 12 2009 Vacuum gauges VAP 5 DVR...

Страница 2: ...controller VACUUBRAND vacuum gauges are the result of many years of experience in construction and practical operation of these vacuum gauges combined with the latest results in material and manufact...

Страница 3: ...his product manual no 999039 16 12 2009 Attention Important notes Not permitted Misuse may cause damage Caution Hot surface Isolate equipment from mains Note Contents Safety information 4 Technical d...

Страница 4: ...ination withVACUUBRAND genuine accessories Obey all other relevant safety requirements e g accident prevention regulations for working in laboratories and explosion protection guidelines and provide p...

Страница 5: ...sary is approximately decontaminated before starting maintenance Use only genuine spare parts and accessories Otherwise safety and performance of the equipment as well as the electromagnetic compatibi...

Страница 6: ...0 1 n i h t i w e u l a v d e t a c i d n i f o 0 1 r a b m 0 0 1 2 r a b m e g n a r t i g i d 1 r a b m 1 t n e i c i f f e o c e r u t a r e p m e T d e i f i c e p s t o n K r a b m 7 0 0 n o i t...

Страница 7: ...peration Display Rear side key button for digital adjustment and electronic scale conversion pressure unit at the side of the housing male connector for mains cable thread M8 for stand rod female conn...

Страница 8: ...yed with a resolution of 0 1 mbar in the range of 0 1 to 10 mbar and with a resolution of 1 mbar for pressures 10 mbar Preparing for operation Connect gauge head with connecting cable to the female co...

Страница 9: ...ace the actual pressure value can be read out The single abbreviations of a command are separated by underscores ASCII 5FH The string is terminated with CR LF ASCII 0DH ASCII 0AH r e t e m a r a p e c...

Страница 10: ...d a e h e g u a G d e t a n i m a t n o c r o r e t l i f e v i t c e t o r p n a e l C y r a s s e c e n f i e n o w e n e s u d n a d a e h e g u a g n a e l C t s u j d a e r g n o r w s i t n e m...

Страница 11: ...h the pumped media see Technical data If the vacuum system is contaminated oil particles etc Replace the protective filter cat no 63 87 29 in the gauge head VSP 5 if necessary Using small screw driver...

Страница 12: ...not be adjusted under vacuum replace gauge head Push key button at the side of the housing with a pencil tip or a small screwdriver Status indication readjustment Set is displayed Disconnect leads fro...

Страница 13: ...ey button to adjust the display to the reference pressure at the vacuum line within the range of 0 10 mbar Note If the nominal ultimate vacuum of a diaphragm pump is used as reference vacuum the accur...

Страница 14: ...ry product returned for repair We submit quotations only on request and always at the customer s expense If an order is given the costs incurred are offset from the costs for repair or from the pur ch...

Страница 15: ...for the returned product that neither toxic corrosive biologically active explosive radio active nor contamination dangerous in any way has occurred the product is free of dangerous substances the oil...

Страница 16: ...agreement of the manufacturer Par la pr sente nous d clarons que le dispositif d sign ci dessus est conforme aux prescriptions de base des directives EU applicables et indiqu s en ci que concerne con...

Страница 17: ...obtained under test conditions for specific applications depend on a number of factors beyond our control It is therefore strictly the users responsibility to very carefully check the validity of app...

Отзывы: