77
20901059_EN_VACUU·SELECT_V1.3_280119
Troubleshooting
Error
`
Possible cause
Remedy
Personnel
Transfer failed
`
`
No USB flash drive
connected.
`
`
Not enough storage
space on the USB
flash drive.
`
Connect a USB
flash drive with
sufficient storage
space.
Specialist
Venting valve
does not oper
‑
ate
`
`
No voltage applied.
`
`
VACUU·BUS plug‑in
connection or cables
defective or not con
‑
nected.
`
`
Venting valve dirty
(polluted).
`
`
Venting valve in sen
‑
sor defective.
`
`
Venting valve deacti
‑
vated.
`
Check VACUU·BUS
plug‑in connection
and cables to the
controller.
`
Clean venting valve.
`
If necessary, use
another external
venting valve.
`
Activate venting
valve in the control
‑
ler.
Specialist