39
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Installation and connection
4.3.4 Venting connection
DANGER
Risk of explosion due to venting with air.
Depending on the application, venting can cause explo
-
sive mixtures to form or other hazardous situations to
arise.
>
>
Never vent processes with air which could form an
explosive mixture.
>
>
In the case of flammable substances, use only
inert gas for venting, e.g., nitrogen (max. 1,2 bar/
17.5 psi abs.).
Venting with ambient air
1
(k)
For venting with ambient air, nothing must be connected to the
VENT port
(k)
.
1 Only valid for sensors with integrated venting valve.
Position VENT port