35
20901101_EN_PC5xx-6xx_select_Serie_V1.3_260819
Installation and connection
4.3 Connection
All condensers of the pumping unit series have a vacuum connec
-
tion and an exhaust gas connection. Perform the connection for
your pumping unit as described in the examples below.
4.3.1 Vacuum connection (IN)
CAUTION
Flexible vacuum hoses can contract during
evacuation.
Connected components that are not secured can cause
injury or damage due to jerky movement (shrinkage) of the
flexible vacuum hose. The vacuum hose can come loose.
>
>
Fix the vacuum hose to the connections.
>
>
Secure connected components.
>
>
Take the maximum shrinkage into account when sizing
the flexible vacuum hose.
NOTICE
Foreign bodies in the suction line can damage the vacuum
pump.
>
Prevent particles, liquids or contaminants from being aspi
-
rated or being able to flow back.
>
Use a sufficiently stable vacuum hose that is designed for
the required vacuum range.
>
Keep hose lines as short as possible.
>
Connect hose lines in a gas‑tight manner to the vacuum
pump.
Observe the following points for optimum results:
>
>
Keep the vacuum line as short as you can with as large a
cross-section as possible.
IMPORTANT!