page 24 of 49
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999100 / 06/05/2008
To chose a pressure value for automatic switching off is
only possible if automatic switching off is activated (in mode
vacuum control with automatic).
➨
Press key MODE simultaneous with
▼
(arrow down).
☞
When the key is pressed shortly, the current pressure for
automatic switching off is displayed for one second (”AE”, i.
e. automatic switching off is preselected at the factory). With
a second tip within one second or continuous pressing:
➨
Select pressure for automatic switching off by using keys
▲
or
▼
(pressure selectable in a range from 1 to 1060 mbar or
”AE”; ”AE”, i. e. automatic switching off is active if a pres-
sure value < 1mbar is selected).
☞
The pressure is stored as new pressure for automatic switch-
ing off when the key is released.
Automatic switching off
MODE
30
mbar
Changing to mode continuous pumping
➨
Press key MODE simultaneous with
▲
(arrow up).
☞
Switching to continuous pumping is only possible, if vacuum
control is not activated (after pressing key STOP).
☞
Adjust the pumping speed choosing an appropriate speed.
➨
Press key MODE again to switch to vacuum control.
☞
Switching to vacuum control is only possible, if pumping
down is activated (after pressing key START).
MODE