If the automatic mode has been preselected
“End”
is displayed.
☞
Switching between automatic switching off activated or not
activated is possible by pressing key
▲
or
▼
.
➨
Confirm by pressing key START/STOP.
☞
The controller switches to normal operation mode (process
control not active).
☞
Automatic switching off as soon as the controller detected
that evaporation is completely finished. The pump is switched
off, the coolant is running on, if a coolant valve is prese-
lected.
☞
Without automatic switching off, the pressure in the vacuum
system is adjusted after evaporation is completed if neces-
sary until the ultimate vacuum of the system.