page 27 of 80
Gas inlet temperatures
Type
ME 2C NT
ME 4C NT
ME 4C NT + 2 AK
Maximum pumping
speed* 50/60 Hz
(ISO 21360)
cfm
(m
3
/h)
1.2 / 1.4
(2.1 / 2.4)
2.3 / 2.6
(3.9 / 4.3)
Ultimate vacuum
(absolute)
Torr
(mbar)
52.5
(70)
52.5
(70)
Rated motor power
hp (kW)
0.24 (0.18)
Maximum permissible range
of supply voltage ( ±10% )
Attention: Observe specifications
of rating plate!
100-115 V~ 50/60 Hz, 120V~ 60 Hz
230 V~ 50/60 Hz
Dual voltage motor
100-115 V~ 50/60 Hz, 120~ V 60 Hz /
200-230 V~ 50/60 Hz
Maximum rated current at:
100-115 V~ 50/60 Hz, A
120 V~ 60 Hz
200-230 V~ 50/60 Hz A
230 V~ 50/60 Hz A
3.4
1.8
1.8
Inlet
hose nozzle for tubing I.D. 3/8" or small flange
KF 16 (hose nozzle DN 10 mm or small flange
KF 16)
Outlet
hose nozzle for tubing I.D. 3/8”
(hose nozzle DN 10 mm)
Dimensions L x W x H
approx.
design 2 AK
in
(mm)
in
(mm)
9.6 x 8.3 x 7.8
(243 x 211 x 198
-
(-)
10.0 x 9.6 x 7.8
(254 x 243 x 198)
12.4 x 9.6 x 11.5
(316 x 243 x 291)
Weight approx.
design 2 AK
lbs. (kg)
lbs. (kg)
22.5 (10.2)
-
24.3 (11.1)
30.0 (13.6)
* Pumping speed of diaphragm pump
Operating condition
Inlet pressure
Permitted range of gas
temperatures at inlet
Continuous operation
> 75 Torr (100 mbar)
(high gas load)
➨
50 °F to 104 °F
(+10°C to +40°C)
Continuous operation
< 75 Torr (100 mbar)
(low gas load)
➨
32 °F to 140 °F*
(0°C to +60°C*)
Short-time
(< 5 minutes)
< 75 Torr (100 mbar)
(low gas load)
➨
14 °F to 176 °F*
(-10°C to +80°C*)
* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)
Содержание MZ 2C NT+2AK
Страница 32: ...page 32 of 80 ME 4C NT 1 5 4 7 6 8 2 ME 2C NT 1 5 4 7 6 8 2...
Страница 34: ...page 34 of 80 ME 8C NT MD 4C NT 1 5 3 4 7 6 2 8 8 5 4 7 6 1 2...
Страница 57: ...page 57 of 80 Fittings and tubing of the different pump models ME 4C NT MZ 2C NT ME 8C NT MD 4C NT...
Страница 73: ...page 73 of 80 Declarations and certificates...
Страница 74: ...page 74 of 80...
Страница 78: ...page 78 of 80...
Страница 79: ...page 79 of 80...