Position clamping bracket (S) with counter-
sunk bores facing upwards.
➨
Align the countersunk bores with the thread-
ed pegs.
➨
Loosely fasten the countersunk screws and
correct the alignment of the valve heads if
necessary.
➨
Tighten countersunk screws with Torx screwdriver T20.
+
Torque: 2.2 ft
.
lb
f
(3 Nm).
➨
Bring the diaphragms (F) into a position, in
which they are in contact with the housing
(C) and centered with respect to the bore.
➨
Put on head cover (J) with valve heads (O)
and connections attached.
+
Pay attention to the correct orientation of
the head covers:
Housing with head alignment pin: The head
alignment pin (A) at the pump housing (C)
has to fit into the recess at the head cover
(J).
Housing with mark (A): Align the recess at
the head cover with the mark at the pump
housing.
➨
Loosely screw in the Allen head screws (H)
at the head covers diagonally at first slightly
with a 5 mm wide Allen key, then tighten.
+
Recommended torque: 8.9 ft
.
lb
f
(12 Nm).
➨
Slide the caps (I) into the head cover.
Valve head with gas ballast or hose nozzle connection:
➨
Insert square nut (K) in the groove of the head cover (J) or position
square nut in the groove and then screw on connection fastener.
+
Loosely fasten fillister head screw (L).
S
H
Содержание MZ 2C NT+2AK
Страница 32: ...page 32 of 80 ME 4C NT 1 5 4 7 6 8 2 ME 2C NT 1 5 4 7 6 8 2...
Страница 34: ...page 34 of 80 ME 8C NT MD 4C NT 1 5 3 4 7 6 2 8 8 5 4 7 6 1 2...
Страница 57: ...page 57 of 80 Fittings and tubing of the different pump models ME 4C NT MZ 2C NT ME 8C NT MD 4C NT...
Страница 73: ...page 73 of 80 Declarations and certificates...
Страница 74: ...page 74 of 80...
Страница 78: ...page 78 of 80...
Страница 79: ...page 79 of 80...