Installing a pump in a vacuum system

➨ 

If  dangerous  or  polluting  fluids  could  be  released  at 

the outlet, install an appropriate system to catch and 

dispose of those fluids.

 Connect a gas-tight exhaust line at the pump outlet if 

necessary. Always vent exhaust gases appropriately 

(e.g., into a fume hood).

 Never block the gas outlet. The exhaust line must al-

ways be free of obstructions (no back pressure) to en-

sure an unimpeded discharge of gas. The cross-sec-

tion of the outlet tubing must be at least the size of the 

pump’s exhaust connection.

 Maximum permissible pressure at the pressure trans-

ducer: 21.8 psi (1.5 bar) absolute.

☞ 

Attention

: At pressures above approximately 810 Torr 

(1080 mbar) the pressure reading becomes incor-

rect. 

The display will flash and a warning message will 

be displayed. Release pressure immediately! Risk of 

bursting!

☞ 

Particles and dust must not be aspirated. If necessary, 

you must install appropriate filters. You must ensure 

their  suitability  concerning  gas  flow,  chemical  resis

-

tance and resistance to clogging prior to use.

☞ 

Make sure ventilation is adequate, especially if the pump 

is installed in an enclosure, or if the ambient temperature 

is elevated. Provide external ventilation, if necessary.

Use and operation

Apart from a short description of the VACUU•SELECT

®

 operating panel, 

this manual describes the mechanical structure of the diaphragm pump.

Operation of the 

VACUU•SELECT

®

 vacuum controller and its func-

tions

 are described in the separate 

VACUU•SELECT

®

 manual

.

Содержание ME 4C VARIO select

Страница 1: ...Instructions for use OI no 20901103 Original instructions EN ME 4C VARIO select MZ 2C VARIO select MD 4C VARIO select Speed controlled Chemistry diaphragm pumps...

Страница 2: ...d intensively including an endurance run of 14 hours Any faults even those which oc cur rarely are identified and can be eliminated immediately After completion of the endurance run every pump is test...

Страница 3: ...required measures must be applied Safety instructions for vacuum equipment FR Attention Le mode d emploi pr sent n est pas disponible dans toutes les langues d Union Europ enne L utilisateur ne doit m...

Страница 4: ...hi seadet k sitseda kui ta ei saa kasutusjuhendist aru Sel juhul peab saadaval olema kogu kasutusjuhendi tehniliselt korrektne t lge Enne seadme kasutamist tu leb kogu juhend l bi lugeda see peab olem...

Страница 5: ...nte letto e compreso prima di operare con lo strumento e devono essere applicati tutti gli accorgimenti richiesti Istruzioni di sicurezza per apparecchi a vuoto JP KR LT D mesio is vadovas n ra pateik...

Страница 6: ...o ci przeczytana i zrozumiana przed podj ciem pracy z urz dzeniem oraz nale y podj wszystkie niezb dne kroki zwi zane z prawid owym uzytkowaniem Wskaz wki bezpiecze stwa do urz dze pr niowych PT Aten...

Страница 7: ...pravljene pa moraja biti tudi vse potrebne meritve Varnostni nasveti za vakuumske naprave SV Varning Denna instruktion r inte tillg nglig p alla spr k inom EU Anv ndaren f r inte starta utrustningen o...

Страница 8: ...al Read the Safety information for vacuum equipment and observe the instructions contained therein Apart from a short description of the VACUU SELECT operating pan el this manual describes the mechani...

Страница 9: ...arts 30 Abbreviations 30 Pump parts 31 VACUU SELECT parts 34 Use and operation 37 Installing a pump in a vacuum system 37 VACUU SELECT Controller 40 Important notes regarding the use of gas ballast 45...

Страница 10: ...page 10 of 76...

Страница 11: ...g from the pump and the pumped substances VACUUBRAND disclaims any liability for inappropri ate use of these pumps and for damage from failure to follow instructions contained in this manual This manu...

Страница 12: ...jury Caution Hot surface General danger sign Disconnect equipment from AC power Electronic components must not be disposed of in the domestic waste at the end of their service life Used electronic dev...

Страница 13: ...any and all warranty claims for those deficiencies Replace the protective caps if the equipment is not used immediately Store the equipment in dry and non corrosive conditions see also Technical data...

Страница 14: ...pends on several parameters like inlet pressure and ambient tempera ture see Technical data pg 26 Particles and dust must not enter the pump Do not pump liquids Ensure that the pump is chemically resi...

Страница 15: ...t with the valve closed to avoid a risk of bursting Maximum permissible pressure at the pressure trans ducer 21 8 psi 1 5 bar absolute Keep the electrical power cord away from heated sur faces Keep th...

Страница 16: ...ng walls etc or else install an external auto matic ventilation system Check fan regularly for dust dirt Clean fan guard grill if necessary to avoid a reduction of ventilation The mains plug is a disc...

Страница 17: ...installation in higher altitudes risk of insufficient cooling conductive pollu tion or external condensation on the pump Do not operate this product near flames To the best of our knowledge the equip...

Страница 18: ...y when being ex posed to elevated temperatures without air substances subject to auto ignition substances which are inflammable without air explosive substances The pumps are not approved for operatio...

Страница 19: ...y emission of such substances from the pump outlet install an appropriate collecting and disposal system and take protective action for pump and environment You must take suitable precautions to preve...

Страница 20: ...to power failure failure of connected components or of parts of the sup ply or change in parameters must not be allowed to lead to dangerous conditions In case of a diaphragm failure or in case of a l...

Страница 21: ...n Provide appropriate safety measures Check prior to starting the process whether the option Autostart menu Settings Basic Settings is enabled Pay attention to the safety symbol hot surfaces on the eq...

Страница 22: ...d repair of vacuum pumps For details and for the online Instructions for repair man ual see www vacuubrand com In normal use the lifetime of the diaphragms and valves is typically 15 000 operating hou...

Страница 23: ...ly moistened cloth To moisten the cloth we recommend water or mild soap Interventions on the device Work on the equipment may be performed only by a qualified person Work on electrical equipment in pa...

Страница 24: ...ed by gases vapors or mists normally don t occur or if they do occur are likely to do so only infrequently and for a short period only Equipment in this category ensures the requisite level of protect...

Страница 25: ...s or Techni cal data After any intervention at the equipment e g repair maintenance the ul timate vacuum of the pump has to be checked Only if the pump achieves its specified ultimate vacuum is the pu...

Страница 26: ...ermissible ambient temperature storage operation F C 14 to 140 50 to 104 10 to 60 10 to 40 Permissible relative atmospheric mois ture during operation no condensation 30 to 85 Maximum permissible inst...

Страница 27: ...vacuum with exhaust tube at outlet Type ME 4C VARIO select MZ 2C VARIO select MD 4C VARIO select Maximum pumping speed 50 60 Hz ISO 21360 cfm m3 h 2 9 4 9 1 66 2 8 2 7 4 6 Ultimate vacuum abso lute wi...

Страница 28: ...periods less than 5 minutes up to 176 F 80 C Ambient temperature range operation 50 F to 104 F 10 C to 40 C Ambient temperature range storage 14 F to 140 F 10 C to 60 C Permissible relative atmospheri...

Страница 29: ...e RJ45 USB port 2x USB A 2 0 max 0 5 A per port Dimensions controller L x W x H without stand 6 0 x 5 0 x 1 6 152 mm x 127 mm x 41 mm Dimensions sensor with small flange with hose nozzle with hose con...

Страница 30: ...amic gold coated if applicable Seal at sensor Chemically resistant fluoroelastomer Measurement chamber PPS Small flange measurement connection PP O ring in small flange sensor FPM Hose nozzle measurem...

Страница 31: ...Diaphragm pump VARIO select 2 Fan 3 Recessed grip 4 Pump rating plate 5 Inlet 6 Gas ballast valve 7 VACUU SELECT controller 8 Mains connection 9 ON OFF switch 10 Outlet silencer 11 Fuse holder 12 VAC...

Страница 32: ...page 32 of 76 MZ 2C VARIO select 9 1 4 10 7 2 6 5 12 3 11 8 9 1 4 10 7 2 6 8 5 11 12 3 3 3 2 2 MD 4C VARIO select...

Страница 33: ...page 33 of 76 13 15 14 VACUU SELECT controller operating panel and sensor Component 13 Touchscreen of operating panel 14 Stand extended 15 VACUU SELECT sensor with connection options...

Страница 34: ...er feet 6 3x connection sockets for VACUU BUS components 7 Recess for VACUU SELECT Sensor 8 Power supply via VACUU BUS connection VARIO pump 9 Rating plate 10 Stand foldable 11 USB port type A 12 RJ45...

Страница 35: ...page 35 of 76 3 4 5 1 2 7 8 11 12 6 9 10 Operating panel Top view Front view Rear view...

Страница 36: ...d VACUU SELECT Sensor Side view Top view Component 1 VACUU SELECT Sensor 2 VACUU BUS plug attachment detachable option 3 Venting valve connection 4 Vacuum screw connection 5 Port for VACUU BUS plug at...

Страница 37: ...ressures above approximately 810 Torr 1080 mbar the pressure reading becomes incor rect The display will flash and a warning message will be displayed Release pressure immediately Risk of bursting Par...

Страница 38: ...uate to maintain recommend ed operating temperature Keep a minimum distance of 2 in 5 cm between the cooling fan and surrounding items e g housing walls etc or else install an external auto matic vent...

Страница 39: ...check the whole system for leaks Secure hose connections at the pump appropriately e g with hose clamps to protect against accidental detach ment To reduce pump noise emanating from the pump exhaust...

Страница 40: ...ration panel Identical com ponents must be configurated beforehand see manual of VACUU SELECT Controller Attention Do not apply off axis forces when assembling or removing plug connections Observe cor...

Страница 41: ...s Therefore carefully choose the position where to connect the sensor Position the VACUU SELECT Sensor and its vacuum line in such a way that condensate cannot flow towards the pressure transducer Con...

Страница 42: ...s for control Button Function Active Locked Start Start Start application only available on the process screen Stop Stop Stop application always possible VENT VENT vent the system option Press button...

Страница 43: ...page 43 of 76 Operating panel Operation see manual of VACUU SELECT Controller Start the pump process Stop the pump process Venting...

Страница 44: ...more than 6500 ft 2000 m above mean sea level check compatibility with applicable safety requirements and adopt suit able measures There is a risk of the motor overheat ing due to insufficient coolin...

Страница 45: ...minutes before re starting the pump Check fan regularly for dust dirt Clean if necessary Avoid overheating e g due to hot process gases A warm up period approximately 15 min is required to en sure tha...

Страница 46: ...ble vapors water vapor sol vents etc The gas ballast valve is open if the arrow on the gas ballast cap is pointing towards the labelling GB With gas ballast valve open the ultimate vacuum will be redu...

Страница 47: ...which may damage the pump materials or form deposits Check and clean pump heads if necessary Has the pressure transducer been exposed to media which may form deposits Clean pressure transducer if nec...

Страница 48: ...ly possible if the pressure is higher than 525 Torr 700 mbar Vent the measurement connection of the vacuum sensor e g VACUU SELECT Sensor and or the vacuum sys tem Make sure that the measurement conne...

Страница 49: ...the measurement connection of the vacuum sensor to a pressure within 0 15 Torr 0 20 mbar Note The accuracy of the value of the reference pressure will directly affect the accuracy of the adjustment If...

Страница 50: ...5154 01 The DAkkS is signatory to the multilateral agreements of the European cooperation for Accreditation EA and of the International Laboratory Ac creditation Cooperation ILAC for the mutual recogn...

Страница 51: ...me Observe all regulations concerning usage and disposal of sol vents Drain the solvent and dispose of in accordance with regulations Repeat cleaning if necessary Rinse the measurement chamber several...

Страница 52: ...pter for gas ballast connection via small flange KF DN 16 20636193 PTFE tubing sold by meter 20638644 VACUU SELECT Sensor 1080 0 1 mbar 20700020 Vacuum gauge VACUU VIEW 1100 0 1 mbar 20635490 without...

Страница 53: ...and com VCL 01 VCL B 10 VCL 02 VCL B 11 VACUU LAN manual flow control module VCL 01 20677106 VACUU LAN shut off manual flow control module VCL 02 20677107 VACUU LAN automatic control module VCL B 10 2...

Страница 54: ...operate VACUU BUS plug in connection or cables defective or not con nected Check VACUU BUS plug in connection and cables to the controller Venting valve dirty Clean venting valve or use another valve...

Страница 55: ...p inlet Vacuum line too long or cross section too small Use a shorter vacuum line with a larger cross section Condensation in the vacuum pump Allow pump to run for some minutes with atmo spheric press...

Страница 56: ...k or diaphragm clamping disc loose Perform maintenance and replace defective parts Ball bearing defective or other than above mentioned causes Contact local distributor Pump stops and does not start a...

Страница 57: ...ith the pro cess chemicals that have been pumped during opera tion Ensure that the pump is decontaminated before maintenance Avoid the release of pollutants Never operate a defective or damaged pump T...

Страница 58: ...agms and the valves must be cleaned and the diaphragms and valves must be checked for cracks or other damage All bearings are encapsulated and are filled with long life lubricant Under normal operatin...

Страница 59: ...th the assembled valve heads and fittings To maintain the diaphragms the valve heads and the fittings need not be disassembled The head covers can be removed along with the as sembled valve heads and...

Страница 60: ...isc with square head screw H Allen screw I Cap J Head cover K Square nut View of the disassembled pump head parts fig MD 4C VARIO select L Fillister head screw M O ring N Valve O Valve head P Hose noz...

Страница 61: ...tely W Opening the hose clip Apply slotted screwdriver as shown and turn Open the hose clip U with a slotted screw driver Pull the tubing off the hose connector Disassemble head covers J to check the...

Страница 62: ...ort disc Lift the diaphragm at the side Place the diaphragm carefully together with diaphragm clamping disc and diaphragm support disc in the dia phragm key Avoid damage of the diaphragm Do not excess...

Страница 63: ...mm wide Note Never use the diaphragm key with any additional tools like tongs or Allen keys without appropriate torque limitation F G Loosen the clamping brackets S on the valve heads O Unscrew at ea...

Страница 64: ...t The valve tongue points at the kidney shaped orifice in the valve seat Outlet side EX Marked with EX next to the valve seat The valve is oriented the same direction as the valve at the inlet side EX...

Страница 65: ...the correct orientation of the head covers Housing with head alignment pin The head alignment pin A at the pump housing C has to fit into the recess at the head cover J Housing with mark A Align the r...

Страница 66: ...is required before the pump achieves its ulti mate vacuum In case of an unusual noise switch off pump immediately and check clamping disc positions If the specified ultimate vacuum is not achieved an...

Страница 67: ...ve fuse by a fuse of the same type see Technical data pg 26 Reassemble holder with fuse to the pump Replacing the device fuse Switch off the pump Disconnect the electrical power cord before un screwin...

Страница 68: ...ed Also decontaminated devices must not returned to VACUUBRAND due to a residual risk The same conditions apply to on site work No repair maintenance return or calibration is possi ble unless the corr...

Страница 69: ...e parts for cosmetic reasons at your request and at your expense Before returning the device Pack the device properly if necessary please order origi nal packaging materials at your costs Mark the pac...

Страница 70: ...rmal wear and tear of wearing parts such as diaphragms seals valves vanes condensers oil and the breakage of glass or ceramic parts for the consequences of chemical electrochemical or electrical influ...

Страница 71: ...ikelnummer Order number Num ro d article 20732450 20732451 20732457 20736550 20736551 20736552 20736557 Seriennummer Serial number Num ro de s rie Siehe Typenschild See rating plate Voir plaque signal...

Страница 72: ...e Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regu lations 2012 S I 2012 No 3032 Diaphragm vacuum pump Type MZ2CVARIOselect MD4CVARIOselect Order number 20732450 2073245...

Страница 73: ...rocedure was performed As defined in GB T 26572 the Maximum Con centration Value limits MCV apply t 0 0 4 0 3 1 0 4 4 4 4 1 4 4 4 0 2 5 0 0 0 2 0 0 0 0 4 0 0 0 2 0 ous environmental pollution cause se...

Страница 74: ...page 74 of 76 4 4 GB T 26572 1 5 0 0 0 2 0 GB T 26572 1 5 0 0 0 1 0 EFUP EFUP 2 1 4 1 2 2 0 1 0 4 0 0 0 2 0 4 0 3 1 0 4 4 4 4 0 0 0 4 4 0 1 4 1 0 2 0 0 0 0 0 2 4 0 3 1 00 4 222 1 00...

Страница 75: ...page 75 of 76...

Страница 76: ...eim 97877 Wertheim GERMANY GERMANY Phone Head office 49 9342 808 0 Sales 49 9342 808 5550 Service 49 9342 808 5660 Fax 49 9342 808 5555 Email info vacuubrand com Web www vacuubrand com Version 2090110...

Отзывы: