page 18 of 56
+
Interruption of the pump (e.g., due to power failure),
failure of connected components or of parts of the sup-
ply, or change in parameters must not be allowed to
lead to dangerous conditions. In case of a diaphragm
failure or in case of a leak in the manifold, pumped
substances might be released into the environment or
into the pump housing or motor.
Comply with all notes regarding proper use of the
pumps, as well as operation and maintenance guid-
ance.
+
The residual
leak rate of the equipment
might render
possible an exchange of gas, albeit extremely slight,
between the environment and the vacuum system.
Adopt suitable measures to prevent contamination of
the pumped substances or the environment.
• Ensure that no parts of your clothing, hair or fingers
can be caught or drawn in at the inlet of the pump.
Never insert fingers or drop any other object into the
inlet or outlet.
• Pumping at high inlet pressure may lead to overpres
-
sure at the gas ballast valve. Pumped gases or con-
densate might be expelled if the valve is open. If an in-
ert gas supply is connected to the gas ballast, ensure
that its inlet pipeline is not contaminated.
•
Pay attention to the safety symbol ”hot surfaces” on
the equipment. Hot parts may cause burns if touched.
Adopt suitable measures to prevent any danger aris-
ing from hot surfaces or electric sparks. Ensure that
hot surfaces of the pump do not cause burns. Provide
a suitable contact guard if necessary.
Prevent the backpressure of gases and the backflow of
condensates at the outlet.
NOTICE