Precautions for Use

Before putting items into the equipment, confirm that the temperature in

the storage chamber has reached the set temperature, and then put the
items in batches. Do not put in items for more than 1/3 of the volume of
the chamber at a time to prevent its temperature from rising too much.

The equipment temperature display is the temperature at the temperature

sensor in the storage chamber. When the equipment is initially running,
the displayed temperature is somewhat different from the actual tempera-
ture at the center of the equipment. However, as the equipment enters a
stable state, the displayed temperature will gradually approach the actual
temperature.

The equipment chamber is designed with a testing hole so that the test

line inside the chamber can be led out for test. After leading out the test
lead, make sure to re-plug the testing hole with insulation material. Other-
wise, the temperature inside the chamber may not reach the temperature
setting and cause condensation around the outside of the through-hole.

Please use diluted neutral cleaner to clean the equipment. Do not use

brushes, acid, gasoline, soap powder, polish or hot water to clean this
equipment, otherwise it may damage the painted surface and plastic rub-
ber parts. Special care must be taken not to use volatile solvents such as
gasoline to wipe plastic and rubber parts.

Cut off the power when the equipment is not used for a long time, and turn

off the battery at the same time.

Try to reduce the opening time of the door for every storage and retrieval

of items to avoid large fluctuations in temperature and humidity inside the
chamber.

After the door is opened, the temperature inside the chamber will rise

sharply for a short time, which is normal, and will recover within one hour
after the door is closed.

7

Содержание VSC350P

Страница 1: ...uct please read this manual carefully and keep it for future reference The design and speci cations are subject to change without prior notice for product improvement Consult with your dealer or manuf...

Страница 2: ...roduct Installation Product Components and Overview User manual Fault Alarm Codes Maintenance e c i v r e S r i a p e R d n a g n i t o o h s e l b u o r T Speci cation Packing List Electrical Schemat...

Страница 3: ...nd Refrigerant and cyclopentane foaming material used for the appliance are flammable Therefore when the appliance is scrapped it shall be kept away from any fire source and be recovered by a special...

Страница 4: ...ms The keys of this series of products need to be properly stored to avoid accidents when children get them and open the door to play Do not connect the zero wire N terminal and the ground wire E term...

Страница 5: ...technology ensures that the tem perature di erence in the chamber is within 2 C International famous brand compressor ensures energy saving environ mental protection high e ciency and low noise equipm...

Страница 6: ...vel or the combined refrigerator freezer is placed in an improper location it may fall down and cause injuries If the power cord needs to be extended the cross sectional area of the exten sion cord sh...

Страница 7: ...power cord If the combined refrigerator freezer is to be stored unused in an unattend ed area for a long period ensure that children do not have access to the combined refrigerator freezer The dispos...

Страница 8: ...e objects with moving parts CAUTION Keep the ULT freezer clear of obstruction around it to ensure smooth ventila tion The combined refrigerator freezer settings shall be checked when the com bined ref...

Страница 9: ...ter leading out the test lead make sure to re plug the testing hole with insulation material Other wise the temperature inside the chamber may not reach the temperature setting and cause condensation...

Страница 10: ...e repair and cause damage to the stored items The ground for installation must be solid and at non combustible and able to withstand the weight of the equipment during operation Good ventilation no di...

Страница 11: ...id su ocation accidents First Power up Follow the steps below when using the equipment for the rst time 1 After the equipment is placed and leveled and cleaned it shall be left for more than 24 hours...

Страница 12: ...The temperature range inside the refrigerator chamber is 2 8 C and the temperature range inside the freezer chamber is 10 25 C Please con rm that the stored items match the environment inside the cham...

Страница 13: ...nd di erent models the actual product may be di erent from the schematic diagrams The actual con guration will de pend on the physical product The schematic diagrams are only for the description of fu...

Страница 14: ...es out if the door is closed The user can also manually turn on the illuminating lamp at this time the illuminating lamp is not controlled by the opening and closing of the door and will always be lit...

Страница 15: ...seconds CAUTION USB ON and END will not be displayed when the data is small USB data manual export When the button is not locked the USB ash drive is con nected and the le is not in the process of gen...

Страница 16: ...When the digital tube gives an alarm and ashes LoF the recorder is not started press and hold and for 3 seconds at the same time LoF disappears and the recorder is started Check Max Min values The MAX...

Страница 17: ...d 5 CLR 0 1 1 0 Clear the minimum and maximum tempera ture value in the current time period 1 Do not clear the minimum and maximum temperature value in the current time period 6 SEt1 0 10 5 Set temper...

Страница 18: ...ing alarm of refrigerator chamber door2 Door opening alarm of freezer chamber PF Power o alarm BL Low battery alarm EE Display board communication fault E1 Temperature display sensor alarm of refriger...

Страница 19: ...body is dirty remove the dirt with a cleaning cloth soaked in neutral detergent and wipe o the residual detergent with a damp cloth then wipe it with a dry cloth Do not pour water on the shell or ins...

Страница 20: ...een stored items Whether there direct sunlight or other heat source radia tion Whether the door opened frequently Whether the environmental temperature too high Whether the air duct blocked Excessive...

Страница 21: ...opentane Shelf Refrigerator chamber 3 shelves Freezer chamber 3 drawers Cooling mode Refrigerator chamber air cooling cycle Freezer chamber direct cooling Outer inner plate PCM sprayed steel sheet Con...

Страница 22: ...al Schematic Diagram Refrigerator fan LED lamp Display control panel USB board Ambient temperature sensor Compressor Compressor Alarm terminal block Normally closed Common Normal open Refrigerator doo...

Отзывы: