Menu
Menu Description
Setting Range
Remarks
n
Set the time of recorder module - year
10
~
50
22 represents 2022
y
Set the time of recorder module -
month
1
~
12
--
r
Set the time of recorder module - day
01
~
31
--
S
Set the time of recorder module - hour
00
~
23
--
F
Set the time of recorder module - min-
ute
00
~
59
--
Pt
Print interval
0
~
240
20min
SCY
Temperature data recording period
0
~
240
0: Shielded recorder
10min
Press the button
to switch parameters n-y-r-s-F-Pt-SCY
Press the button
or the button
to set the parameter value. After setting, press
and hold the button
to save the parameter.
CAUTION: Modifying the time and logging interval will cause the recorder data to be
cleared and the data recorder to be restarted with the digital tube displaying "LoF". To
eliminate it, see Operations to Start Recorder: When the digital tube gives an alarm
and flashes "LoF", the recorder is not started; press and hold
and
for 3 seconds
at the same time, "LoF" disappears and the recorder is started.
Check Max/Min values
The MAX1, MAX2, MIN1 and MIN2 parameters of the user menu show the maximum
and minimum values of the temperature during the current time period
Battery voltage check
Press
+
, the digital tube of the freezer chamber shows the battery voltage
Function Settings
User Parameter Settings
Unlock:
In normal operation, press
and
for 3 seconds at the same
time, the upper chamber digital tube displays "PS1", the lower chamber digital
tube displays the parameter code "0000". Press the button
and enter the
password "0005". " (When entering the user menu password, enter "0099" to
restore the button lock password to the default "0005".) , it is unlocked at this
time. After unlocking, press the button
for 3 seconds, the digital tube will
display the parameter code "MAX" and the setting and adjusting parameters
are entered.
14