6
2.3
Door
Space Requirements
The unit's door must be able to open fully as shown.
W
×
D
≥
1040mm*1030mm
2.4
Levelling
the Unit
To do this adjust the two levelling feet at the front of the unit.
If the unit is not level, the door and magnetic seal alignments
will not be covered properly.
2.5
Cleaning
Before Use
Wipe the inside of the unit with a weak solution of bicarbonate soda. Then rinse with warm water using a 'Wrung-
out' sponge or cloth. Wash the shelves and salad bin in warm soapy water and dry completely before replacing in
the unit. Clean the exterior of the unit with a damp cloth.
If you require more information refer to the Cleaning section.
2.6
Before Using Your Unit
When using the device for the first time, please follow these steps:
1.Under no-
load condition, connect the power cord to a special socket with appropriate specifications.
2.After turning on the power supply, turn on the battery switch of the device, check if the illumination lamp in the
cabinet is under normal operation.
3.Set recorder time, data recording interval time, etc according to actual need, please refer to the operation
instructions below for details.
4.Set parameter of cabinet temperature, high(low) temperature alarm deviation, etc according to actual need.
Default temperature setting is 5
°
C, high temperature alarm is 8
°
C and the low temperature alarm is 2
°
C, please
refer to the operation instructions below for details.
5.
“
H1
”
alarm occurs at the initial operation stage, indicates that the temperature in cabinet is high. The alarm
will stop when the temperature in cabinet drops to the set temperature.
6.
“
LoF
”
alarm occurs at the initial operation stage, indicates that the data recorder is not started. Please start the
data recorder according to the operation instructions below.
7.
“
BL
”
alarm occurs at the initial operation stage, indicates low battery power alarm, and alarm will be
automatically relieved after 2-6 hours of power on.
8.Observe the equipment operation for more than3 hours, ensure the performance is normal, store items in
cabinet in batches.
2.7
Interior Accessories
Multi layer encryption shelf design and adjustable height can maximize the use of space, The shelf bearing capacity
is 15Kg.
The inner LED lights, bright and energy-saving, make the stuff in the box clear at a glance
W
D
load condition, connect the power plug to a special socket with appropriate specifications.