2. INSTALLTION

2.1

 

Handling 

1)Protect the Pharmacy

Refrigerator in moving it,Same as shown as left photo, please move it by handcart with cushion

2)Remove all packing materials and bottom cushion, the  move into house for placement
3)After moving it to appropriate location, wait for 2 hours before power on.

2.2

 

Location 

When  selecting  a  position  for  your  unit  you  should  make  sure  the  floor  is  flat  and  firm,  and  the  room  is  well 
ventilated. Avoid locating your unit near a heat source or radiator. Also avoid direct sunlight as it may Affect the 
operation  of  the  whole  machine.  Extreme  cold  ambient  temperatures  may  also  cause  the  unit  not  to  perform 
properly. This unit is not designed for use in a garage or outdoor installation. Do not drape the unit with any covering.

 

When installing the unit, ensure that 20 cm of free space is left at both sides, 15 cm at the rear and 20 cm at the top 
of the unit. This will allow cold air to circulate around the Pharmacy Refrigerator and improve the efficiency of the 
cooling process.   
The  Pharmacy  Refrigerator  is  only  for  indoor  use.  Ambient  temperature  is  within 16°C-32°C,  and recommended 
temperature is within 18°C-25°C. If necessary, air conditioning system should be used. 
Environmental humidity: lower than 80%Rh. 
Avoid heavy dust. 
Avoid mechanical sway or vibration. 
The altitude of working position of Pharmacy Refrigerator is lower than 2000 m. 
Input voltage: between (220~240)±10% V. 

Ingress protection

IPX0 

Over Voltage Category 

 

Pollution Degree 2 
The Pharmacy Refrigerator is sensitive to ambient temperature. When installed in other environments than the one 
aforesaid, the Pharmacy Refrigerator can't operate normally. Please improve the environment before use. 
It is prohibited to install the Pharmacy Refrigerator in an open environment. When the Pharmacy Refrigerator is wet 
by rain, it may cause electric leakage or electric shock. 

Содержание VS50

Страница 1: ...keep it for future reference The design and speci cations are subject to change without prior notice for product improvement Consult with your dealer or manufacturer for details PHARMACY REFRIGERATOR...

Страница 2: ...ries 6 3 PRODUCT OVERVIEW 7 4 TECHNICAL SPEC 7 5 OPERATION 8 5 1 Opera on and display panel 8 5 2 Key opera on 9 5 3 Noise Inside the Unit 13 5 4 ALARM 13 6 CLEANING 14 6 1 Cleaning the Interior and t...

Страница 3: ...building in clear of obstruction WARNING Do not use mechanical devices or other means to accelerate the defrosting process other than those recommended by the manufacturer WARNING Do not damage the r...

Страница 4: ...r more must be used and copper core conductors in the wall connected to the power socket must also have a cross sectional area of 4mm2 or more This series of products are strictly prohibited to put in...

Страница 5: ...socket are plugged firmly and reliably otherwise it may lead to fires When unplugging from the power socket hold the power plug firmly and remove it Do not pull the wires of the power plug directly If...

Страница 6: ...he insulation and cause leakage or electric shock Do not insert metal objects such as nails or wires into any openings and gaps of the Pharmacy Refrigerator or into any vent used for internal air circ...

Страница 7: ...at both sides 15 cm at the rear and 20 cm at the top of the unit This will allow cold air to circulate around the Pharmacy Refrigerator and improve the efficiency of the cooling process The Pharmacy R...

Страница 8: ...to the operation instructions below for details 4 Set parameter of cabinet temperature high low temperature alarm deviation etc according to actual need Default temperature setting is 5 C high temper...

Страница 9: ...Model number VS50 External dimensions W D H mm 485 450 550 Effective volume 42L Door body Transparent glass door Foaming agent Cyclopentane Cooling mode Air cooling circulation Shell plate Inner Plat...

Страница 10: ...on Button Locked off Button unlocked Power failure on Power off off Power on Printer disabled on Printer working off Printer not working Door heat disabled on Door heat working off Door heat not worki...

Страница 11: ...locked if there is no operation for 60s or Up key and Down key are pressed for 3s at the same time b When the key is locked only ambient temperature can be checked and Mute key and Light switch key c...

Страница 12: ...Lower deviation on low temperature alarm 0 0 10 0 0 cancel the alarm 3 n Check the current date year y Check the current date month r Check the current date day S Check the current date hour F Check t...

Страница 13: ...o 1 Every time the door is opened and closed it will be heated for T9 minutes If the door is opened and closed again during the heating period the heating time will be reset Door heating mode P1 is se...

Страница 14: ...eps once the digital tube displays the upload data code on in case of large amount of data and the USB flash disk generates the PDF files on data of current month and last month After the completion o...

Страница 15: ...bient temperature alarm value tH1 sensor is normal and not shielded then the buzzer will alarm and the digital tube will flash the high ambient alarm code H2 alternately at the interval of 3 seconds e...

Страница 16: ...fied technician for assistance 7 MAINTENANCE 7 1 Care When Handling Moving Your Unit Hold the unit around its sides or base when moving it Under no circumstances should it be lifted by holding the edg...

Страница 17: ...t heat during operation specially in summer this is caused by the radiation of the condenser and it is a normal phenomenon Condensation condensation phenomenon will be detected on the exterior surface...

Страница 18: ...16 REQUESTING ASSISTANCE OR SERVICE If you have any doubt or extra assistance please call Customer Service Vacc Safe Telephone 1300 459 140...

Отзывы: