Continuous level measuring system
NB 3000
Technical information / Instruction manual
page G
Co
ne h
ei
gh
t C
A
ir d
is
ta
nc
e A
M
ax
. m
ov
e d
is
ta
nc
e M
Si
lo h
ei
gh
t H
100%
0%
Quickset menu
Max. adjustable length of 30/40m depending on ordered version.
* Factory-provided
Max. move distance M
Ensures that the weight does not enter into the silo outlet.
Silo height H
Definition of 0% level output.
Note: If the maximum move distance M is smaller than
the silo height H, the measured value will always be more
than 0%.
Air distance A
Definition of 100% level output.
Cone height C
Enables to set the current output as volume.
C =0 Current output indicates material level
C >0 Current output indicates material volume
Note:
When using the digital pulse output (terminal 5/6/7, see page G18/22) the parameters
silo height H, air distance A and cone height C have no influence on the measurement
value.
The Quickset menu is used for fast and easy start-up of the system.
If the unit is working in normal operation (measurement mode), the SETUP button brings up the Quickset menu.
Press to return to
measurement mode
Measurement mode
0,0m (0,0ft)
Max. move distance M
Silo height H
Air distance A
Cone height C
Language
*Deutsch
English
Francais
Russian
Unit
*Meter
Feet
Programming NB 3000 Nivobob
0..30/40m
(0..100/133ft)
*1m (3ft)
0..30/40m
(0..100/133ft)
*0m (0ft)
0..30/40m
(0..100/133ft)
*0m (0ft)
0..30/40m
(0..100/133ft)
*0m (0ft)
see page
G4-G6
X+