holes on the front of the transilluminator. The cover is adjustable to varying angles 
for access to the filter surface.  
 
If  you are not using the transilluminator with an imaging system darkroom, do not 
operate the unit without securing the cover. If the cover is missing, a UV Blocking 
Faceshield must be worn to avoid UV exposure to the skin. UV Blocking Eyewear 
should be worn even with the cover in place to avoid accidental UV exposure. 
 

 

Using the Transilluminator 

 

Place gel/sample on the filter area. It is recommended to place the gels on a Gel-Tray to 
protect the filter surface from cuts and scratches. It is recommended that gloves be worn 
to avoid contact with gel and staining agents. 
 
Press the ON/OFF switch to ON. The UV tubes within the unit should be glowing beneath 
the filter. If using the transilluminator with an imaging system, the system’s main power is 
required to be in the ON position. 
 
When using a transilluminator with multiple UV wavelengths, dial the knob to the 
appropriate wavelength setting.  
 
When using the Variable Intensity models, use the variable intensity settings as follows: 
 

 

High

: allows for UV excitation of fluorophores on gels for routine photography 

and for excitation of gels with low sample concentration  

 

Medium

: Excellent for viewing and quick single-band excision 

 

Low

: Allows for positioning and preparation of gels, excising multiple bands and 

focusing for photography 

 
After viewing/photographing the sample, turn the transilluminator off. 
 

Cleaning and Care of the Transilluminator 

 
Clean unit surface with a damp soft cloth or sponge.  Never use abrasive cleaners (can 
damage the UV filter surface).  
 
To protect the filter glass and minimize moisture and liquids on the glass, it is 
recommended that you use a UV transmitting Gel-Tray. Refer to the Replacement Parts 
for ordering information. 
 

Replacing Tubes in the Transilluminator 

 
1.  Disconnect the transilluminator for the electrical supply. 

 

2.  Remove the filter cover: Use a Phillips head screwdriver to remove the four screws 

on the sides of the unit. Lift the filter cover off the unit. 

 

3.  Remove the reflectors on the left and right side of the unit: Slide the reflectors up out 

of the unit. 

 

4.  Remove the tube: Carefully rotate the tube and slide out of the socket. Replace with 

a new tube by sliding the tube into the socket and rotating into place. 

 

5.  Insert the reflectors back into place and reattach the filter cover. 
 
 

Содержание TFM-20V

Страница 1: ...P LLC Ultra Violet Products Ltd 2066 W 11th Street Upland CA 91786 Unit 1 Trinity Hall Farm Estate Tel 800 452 6788 909 946 3197 Nuffield Road Cambridge CB4 1TG UK Fax 909 946 3597 Tel 44 0 1223 420022 Fax 44 0 1223 420561 Web site www uvp com 81 0278 01 Rev C ...

Страница 2: ...d electrical consumption Select models include variable intensity with High Medium and Low settings The electronic ballast has a universal voltage capabilities NOTE Though the manual refers to the midrange UV waveband as 302nm others refer to this region as 300nm or 312nm The spectral output of all these regions is the same A Word of Caution UV Transilluminators are powerful sources of UV radiatio...

Страница 3: ...0424 02 302nm 230V 50Hz 25 x 30 cm 4 x 25W Variable TFM 40V 95 0421 01 302nm 100 115V 60Hz 20 x 40 cm 4 x 25W Variable TFM 40V 95 0421 02 302nm 230V 50Hz 20 x 40 cm 4 x 25W Variable The following models have 365nm longwave UV TFL 40V 95 0420 01 365nm 100 115V 60Hz 20 x 40 cm 4 x 25W Variable TFL 40V 95 0420 02 365nm 230V 50Hz 20 x 40 cm 4 x 25W Variable The following models have 254nm shortwave UV...

Страница 4: ... the UV Blocking Cover is not being used UV light may escape through the holes dedicated to accepting the bracket pins of the UV Blocking Cover Remove the black safety plugs from their package Insert the safety plugs through the holes as shown Set Up Place the transilluminator on a level work surface Be sure that an air space exists around the bottom of the work surface This space allows for the p...

Страница 5: ...tensity settings as follows High allows for UV excitation of fluorophores on gels for routine photography and for excitation of gels with low sample concentration Medium Excellent for viewing and quick single band excision Low Allows for positioning and preparation of gels excising multiple bands and focusing for photography After viewing photographing the sample turn the transilluminator off Clea...

Страница 6: ...midrange UV 34 0072 01 Qty 4 Required Tube 25 Watt 365nm longwave UV 34 0060 01 Qty 4 Required Tube 25 Watt 254nm shortwave UV 34 0073 01 Qty 4 Required Ballast electronic 2UV models 42 0068 03 115V 42 0068 05 230V Ballast electronic variable intensity models 42 0067 03 115V 42 0067 05 230V Fuse 3 2A 250V Slo Blo 56 0002 05 Qty 2 Required Fuseholder 5 x 20mm RoHS 56 0023 02 Qty 2 Required Switch 5...

Страница 7: ...o 2 years from date of purchase The foregoing warranty of UVP shall be of no force and effect if buyer has modified or damaged the product Tubes and filters are warranted for 90 days All warranties or merchantability and fitness for any purpose and all other warranties expressed or implied except those expressly set forth herein are deemed waived and excluded UVP s duty under the warranty is limit...

Отзывы: