Introduction 

 

The  UVP  Transilluminator offers the researcher uniform and intense sources of ultraviolet light 
(radiation) in a compact package. The special design emits high intensity excitation UV wavelength 
for back-illumination of transparent fluorescent materials. The 302nm UV back-illumination provides 
a highly sensitive method to detect double-stranded nucleic acids that have been labeled with 
fluorescent dyes such as ethidium bromide or acridine orange. Single stranded nucleic acids may 
be detected, but with a lower excitation wavelength more sensitive for nucleic acid visualization 
than the 365nm model.  
 
High Performance UV Transilluminators are fitted with the exclusive 25 watt UV tubes. The 
transilluminator is uniquely designed with increased UV intensity and uniformity, instant on 
capabilities, no lamp flicker and reduced electrical consumption.  Select models include variable 
intensity with High, Medium and Low settings.  The electronic ballast has a universal voltage 
capabilities. 
 

NOTE:

  Though  the manual refers to the midrange UV waveband as 302nm, others refer to this 

region as 300nm or 312nm.  The spectral output of all these regions is the same. 
 

A Word of Caution:  

UV  Transilluminators are powerful sources of UV radiation that will cause 

damage to unprotected eyes and skin. Before operating any unit, be sure all personnel in the area 
are properly protected. If not using the transilluminator with an imaging system  darkroom,  a UV 
Blocking Cover  should be attached to the transilluminator. Even though this cover blocks the 
ultraviolet radiation emitted by the unit, UV Blocking Eyewear should be worn as well.  
 
UVP UV Transilluminator features: 

 

 
 

 

UV Blocking Cover is 
included with the 
transilluminator. The cover 
is hinged and adjustable to 
varying heights 

Models include a control 
knob for selection of 
wavelength or intensity 

Power switch 

UVP’s long life filters are 
available in a variety of 
filter sizes  

Содержание TFM-20V

Страница 1: ...P LLC Ultra Violet Products Ltd 2066 W 11th Street Upland CA 91786 Unit 1 Trinity Hall Farm Estate Tel 800 452 6788 909 946 3197 Nuffield Road Cambridge CB4 1TG UK Fax 909 946 3597 Tel 44 0 1223 420022 Fax 44 0 1223 420561 Web site www uvp com 81 0278 01 Rev C ...

Страница 2: ...d electrical consumption Select models include variable intensity with High Medium and Low settings The electronic ballast has a universal voltage capabilities NOTE Though the manual refers to the midrange UV waveband as 302nm others refer to this region as 300nm or 312nm The spectral output of all these regions is the same A Word of Caution UV Transilluminators are powerful sources of UV radiatio...

Страница 3: ...0424 02 302nm 230V 50Hz 25 x 30 cm 4 x 25W Variable TFM 40V 95 0421 01 302nm 100 115V 60Hz 20 x 40 cm 4 x 25W Variable TFM 40V 95 0421 02 302nm 230V 50Hz 20 x 40 cm 4 x 25W Variable The following models have 365nm longwave UV TFL 40V 95 0420 01 365nm 100 115V 60Hz 20 x 40 cm 4 x 25W Variable TFL 40V 95 0420 02 365nm 230V 50Hz 20 x 40 cm 4 x 25W Variable The following models have 254nm shortwave UV...

Страница 4: ... the UV Blocking Cover is not being used UV light may escape through the holes dedicated to accepting the bracket pins of the UV Blocking Cover Remove the black safety plugs from their package Insert the safety plugs through the holes as shown Set Up Place the transilluminator on a level work surface Be sure that an air space exists around the bottom of the work surface This space allows for the p...

Страница 5: ...tensity settings as follows High allows for UV excitation of fluorophores on gels for routine photography and for excitation of gels with low sample concentration Medium Excellent for viewing and quick single band excision Low Allows for positioning and preparation of gels excising multiple bands and focusing for photography After viewing photographing the sample turn the transilluminator off Clea...

Страница 6: ...midrange UV 34 0072 01 Qty 4 Required Tube 25 Watt 365nm longwave UV 34 0060 01 Qty 4 Required Tube 25 Watt 254nm shortwave UV 34 0073 01 Qty 4 Required Ballast electronic 2UV models 42 0068 03 115V 42 0068 05 230V Ballast electronic variable intensity models 42 0067 03 115V 42 0067 05 230V Fuse 3 2A 250V Slo Blo 56 0002 05 Qty 2 Required Fuseholder 5 x 20mm RoHS 56 0023 02 Qty 2 Required Switch 5...

Страница 7: ...o 2 years from date of purchase The foregoing warranty of UVP shall be of no force and effect if buyer has modified or damaged the product Tubes and filters are warranted for 90 days All warranties or merchantability and fitness for any purpose and all other warranties expressed or implied except those expressly set forth herein are deemed waived and excluded UVP s duty under the warranty is limit...

Отзывы: