Chapter 1: Introduction
1-6 UT100 Handheld Pulse Oximeter Operation Manual
CAUTION! Do not allow water or any other liquid to spill onto the monitor.
Unplug the external power supply from the monitor before
cleaning or disinfecting the monitor. Evidence that liquid has
been allowed to enter the monitor voids the warranty.
CAUTION! Ensure the device’s AC rating is correct for the AC voltage at
your installation site before using the monitor. The monitor’s AC
rating is shown on the external power supply. If the rating is not
correct, do not use the monitor; contact the UTECH Co., Ltd
after-service department, or your local distributor, for help.
CAUTION! Chemicals used in some cleaning agents may cause brittleness
of plastic parts. Follow cleaning instructions in this manual.
Notes
NOTE! Batteries are user replaceable. Follow local governing ordinances
and recycling
instructions regarding disposal or recycling of device
components, including batteries.
NOTE! When using AC power, the Oximeter is a class II device with
functional earth. This earth connection is for device
electromagnetic compatibility and does not provide protection to
the patient or user.
NOTE! It is recommended that batteries be used with the monitor when
operating with AC power to prevent monitor shutdown with loss of
AC power.
NOTE! All user and patient accessible materials are non-toxic.
NOTE! Each input and output connection of the monitor is electrically
isolated.