7. Date and Time setting
Choose ”Pair” in “Serial” menu to open the date and time setting dialogue box, it will display the system clock of PC,
press “Confirm” or exit the dialogue box directly. After the software send date and time setting data to the scale, it
will display “Pair successful.”.
16.
Calibration
Calibration
Calibration
Calibration
1.
When the calibration switch is ON, turn on the scale, or make the calibration switch ON when the scale is under the
weighing mode, the scale will enter the calibration mode automatically.
2.
After entering into the calibration mode, the WEIGHT window will display CAL.ON, which means the calibration switch
is on. The UNIT PRICE window displays "Unit.0" or "Unit.1", which means the chosen calibration weighing unit is
kg(Unit.0) or lb(Unit.1), the TOTAL PRICE window displays empty.
3.
Use Unit key to choose the calibration weighing unit kg or lb (the corresponding unit indicator will be lightened on),
use ENTER key to confirm the unit and go to the next step.
4.
The WEIGHT window still displays CAL.ON, UNIT PRICE window displays "unLoAd" (this means that the scale is ready
The calibration switch is located underneath
the scale base.
When the switch is set towards the USB or
RS232 interface, it means CAL switch is on.
When the switch is set towards the power
switch, it means the CAL switch is off.