Page 30

TROUBLESHOOTING

Error Codes

Reason

Solution

UUUUUU

1. Overload
2. Wrong connection with 

load cell
3. Load cell has quality 

problem

1. Reduce the weight
2. Check load cell connection 
3. Inspect load cell; Check the
input/output
4. See Q&A section

nnnnnnn

Calibration is no good
Wrong connection with load 

cell Load cell has quality 

problem

1. Make sure scale is level
2. Check load cell connection
3. Check load cell input and 

output
resistance
4. See Q&A section

Err1

Weight is not used during 

calibration, or the weight is 

above the maximum capacity

Use correct weight within the 

defined range

Err2

Weight is below the minimum 

requirement during 

calibration.

The calibration weight 

minimum is 10% of the 

max. capacity set in C04. 

Recommended to use 60%-

80% of max. capacity if 

possible

Err3

The input signal is negative 

during calibration

1. Check all wire connections
2. Check load cell
3. Recalibrate
4. PCB replacement needed if 

steps 1-3 fail

Err4

Unstable signal during 

calibration.

After the platform is stable, 

start calibration

Err5

EEPROM Error

Change PCB

Err6

Exceed Zero Range

See Q&A section

Содержание US-6011 Series

Страница 1: ...US 6011 SERIES INDICATOR USER S MANUAL US 6011SS 4 20MA US 6011SS R usameasurements com...

Страница 2: ...4 PREPARATION SET UP 5 POWER SUPPLY 7 DISPLAY AND KEY DESCRIPTION 8 OPERATING INSTRUCTIONS 9 CALIBRATION 12 INDICATOR PARAMETER SETTINGS 14 CONNECTORS 21 RS232 SERIAL OUTPUT FORMAT 23 US 6011SS 4 20 m...

Страница 3: ...ubset to the zero tracking range it is the time allowed for the scale to fall within the zero tracking range tolerance and still qualify to be zero d Overload Range Weight allowance that is out of the...

Страница 4: ...ed Maximum capacity to prevent damaging the inner load cell 9 Stay with in the unit s rated load limit 10 Do not leap on the scale 11 Only professional staff should conduct the calibration inspection...

Страница 5: ...date Modify gravity based on location 100 manual zero range setting Automatic error warning alarm Splash proof keyboard and display RS 232 Port for optional Scoreboard RS 232 Port for connecting to Pr...

Страница 6: ...pn 6 ppm K Sensitivity internal 0 3 v d Input voltage 30 to 30Mv DC Excitation circuit 5 VDC 4 wire connection 6 load cell of 350ohm max AC power AC 100 250V use only the included 9V adapter supplied...

Страница 7: ...month to keep the battery in its best condition When you want to discharge the battery you are required to leave the indicator on until the indicator powers goes off Remove the battery when it is not...

Страница 8: ...t product ZERO Indicates that the scale is zero PRINT Means data should be printed confirms inputs Shows that the scale is at zero Shows that the scale is stable Gross Shows when in Gross weight mode...

Страница 9: ...e indicator is in gross mode When you do that the current weight on the scale will be tared and the scale will enter the net mode Add your product to the scale to weigh without the pallet s weight If...

Страница 10: ...ing between 2 sets of numbers If you want to print the accumulated total hold the PRINT key for one second while the last 4 digits of the total are shown To exit accumulation mode wait for the last 4...

Страница 11: ...the average weight from those 3 seconds Press HOLD again to exit holding mode Print If the indicator is connected to a printer and the weight on the scale is stable press the PRINT key to print the c...

Страница 12: ...C2 set to 1 the scale will read in 0 1 lb increments 12 Press PRINT to set the value The display will now show C04 13 Press PRINT to access the C4 channel The display will show 14 Enter in the maximum...

Страница 13: ...tightly and verify that the platform is level To Calibrate using only multiple calibration weights Linear Calibration 1 Press PRINT to continue The display will now show C06 2 Press PRINT to access t...

Страница 14: ...g or lb 2 Press and hold the HOLD and PRINT key at the same time for 2 seconds 3 Navigate through the settings C01 to C45 as shown in the table 4 below by using the arrow buttons and return buttons as...

Страница 15: ...ge within 3 minutes 5 turn off display if no change within 5 minutes LCD Version OP900B 0 turn off the backlight 1 backlight only when the weight changes or keyboard is pressed2 constant backlight Hol...

Страница 16: ...adhesive label thermal printer 7 Print to Zebra large adhesive label thermal printer 8 Reserved Baud Rate C19 0 1200 for OP 910 remote display 1 2400 2 4800 3 9600 for all printers and OP 910X OP 910...

Страница 17: ...verload range setting Negative Display C25 0 9d 10 10 max capacity 20 20 max capacity 50 50 max capacity 100 100 max capacity Standstill Time C26 0 quick 1 medium 2 slow Standstill Range C27 d divisio...

Страница 18: ...elay output function 1 2 turn on relay output function 2 3 Reserved menu Gravity of Calibration Location C36 9 7000 9 9999 Gravity of Destination C37 9 7000 9 9999 Version No C38 Print Mode C41 0 auto...

Страница 19: ...Weighing Unit C01 1 Decimal Setting C02 0 Graduation Setting C03 1 Maximum Capacity C04 1000 Zero Calibration C05 0 Calibration C06 0 Restore Default C07 0 Warning Tone C08 1 Automatic Power Off C09 0...

Страница 20: ...ter C29 2 Print Time and Date C30 0 Analog Output Setting C31 1 Calibrate Current C32 4 Relay Output Setting C33 1 Multi connection add C34 0 Wireless Communication C35 6 Wireless Communication C36 9...

Страница 21: ...ods between the load cell and indicator Quick Disconnect as shown below Hardwire Using Inner Terminal Block Connection Note Make sure you follow all the anti static rules to avoid damage to your indic...

Страница 22: ...Negative output signal from load cell EXC EX BLACK Negative excitation voltage to load cell DB9 Connection 9 pin Serial Connector The DB9 9 pin serial connector is used for different purposes dependin...

Страница 23: ...is a list of the serial parameters 8 data bits 1 stop bits No parity No handshaking Note With the RS232 Output option we have data logging software available as seen in Insert figure no Below are the...

Страница 24: ...XXXXXXX 1 1 0 XXXXX X 0 0 1 XXXX XX 1 0 1 XXX XXX Bits3 4 0 1 X1 1 0 X2 Satate B BitsS function Bits0 gross 0 net 1 Bits1 Symbol positive 0 negative 1 Bits2 Overload or under zero 1 Bits3 dynamic 1 Bi...

Страница 25: ...st Mode C18 3 In this mode the indicator can receive ASCII commands as listed below Command Name Function 1 Tare Save and clear tare 2 Zero Zero gross weight 3 Print Print the weight 4 G W N W Read gr...

Страница 26: ...nding Mode C18 5 STX POL XXXXX XX LB KG GR NT STAT CR LF Polarity SP Positive Negative Weight Data Gross Net GR Gross NT Net Line Feed Start Transmission Units lb punt kg kilogram pcs pieces Status SP...

Страница 27: ...the scale s internal 9VDC power supply will need to be on the same ground plane for the output current to be synced and measured correctly Please refere to the figure below as a visual clarification o...

Страница 28: ...d HOLD key to go into configuration mode Go to C32 and press PRINT key The display should show out 4 and output should be at 4mA Press the up down arrow keys and the out will increase decrease Note X...

Страница 29: ...trol function Output overload control signal Out2 Turn on compliance control function Output compliance control signal Out3 Turn on under load control function Output under load control signal Out4 Tu...

Страница 30: ...See Q A section Err1 Weight is not used during calibration or the weight is above the maximum capacity Use correct weight within the defined range Err2 Weight is below the minimum requirement during c...

Страница 31: ...to 100 if possible using the up arrow key 7 If 100 is not available change to 20 8 Press PRINT key to enter your selection 9 The screen will display C21 10 Press PRINT key to enter C21 parameter 11 Ch...

Страница 32: ...You can send an email to sales usameasurements com for any questions about your sales or support Alternatively call 800 711 2237 for sales and support Visit our website at usameasurements com...

Отзывы: