35

When  test  the  sample  which  contains Vitamin  C

may  lead  the  occult  blood  and  glucose  readings

may be lower than the actual value.

9) Drug-administered 

urine 

and 

visual

hematuria can affect test results.

Drug-administered  urine  and  visual hematuria  can

affect test results and may not get the accurate test

results.

3.2 Reagent Strips Usage Precautions

1) Please use the special reagent strips.

Use only the URIT reagent strips and carefully read

through the package inserts of the reagent strips.

2) Please confirm before use.

Check  the  expiration  date  of  reagent  strips  before

use.  Do  not  use  the  expired  strips  or  strips  that

have  discolored  pads  even  if  they  are  within  the

expiration date.

3) Reagent strips should be prepared before the

test.

Only take out the required number of reagent strips

Содержание UC-32A

Страница 1: ......

Страница 2: ...1 2 Intended Use 13 1 3 Contraindication 14 1 4 Specifications 14 1 5 Structure and Principle 17 1 6 Appearance and Components 19 2 Installation 23 2 1 Action upon delivery 23 2 2 Instructions for us...

Страница 3: ...Checkup 47 5 1 Checkup 47 5 2 Checkup Operate 48 6 Function Introduction 50 6 1 Function Introduction of Printer 50 6 2 Test Function 50 6 3 Introduction of Testing Result 51 6 4 Reviewing Results 53...

Страница 4: ...8 Storage and Handling 61 8 1 Handling 61 8 2 Storage 61 9 Troubleshooting 62 Appendix 65...

Страница 5: ......

Страница 6: ...f the manual and reserves the right to renovate the relevant contents without separate notification Some of the demonstration pictures are for reference and subject to real object if any differences A...

Страница 7: ...inal purchase or installation URIT assumes no liability in the following situations even during the period of warranty 1 Failure due to abuse the instrument or neglect the maintenance 2 Use reagents a...

Страница 8: ...hnical service and troubleshooting are provided by URIT If the instrument has malfunction please contact the agency authorized by URIT CAUTION THE INSTRUMENT IS FOR PROFESSIONAL AND PRESCRIPTION USE O...

Страница 9: ...4 Web www urit com Email service uritest com Supplied by URIT Medical Electronic Co Ltd VERSION 12 2020...

Страница 10: ...instrument and follow the instructions presented in the manual to perform measurements Caution This instrument uses urine as sample which may carry pathogenic microbes that can cause infectious diseas...

Страница 11: ...ronic products recycling symbol Environment Friendly Use Period Authorized representative in European Community Class equipment direct current DC Caution Biological risks Manufacturer Serial number Ke...

Страница 12: ...rom the instrument when the battery is used up or the instrument is not used for a long time n Replace the waste batteries please in accordance with the relevant local laws and regulations properly ha...

Страница 13: ...provided or recommended by the manufacturer n Turn off the power switch and unplug the power cord immediately if the instrument gives off odor or smoke otherwise it will cause fire electric shock or...

Страница 14: ...gnetic sources or devices that generate electromagnetic wave n Keep the instrument away from liquids dust and physical shocks or impacts n After the instrument is abandoned it should be handled in acc...

Страница 15: ...e with the relevant local regulations n The result tested by the instrument can only be applied to screening some related diseases in the group according to the detection results of human urine It can...

Страница 16: ...ay conveyor and the instrument will test automatically The test results are displayed on display screen and it can also be sent to external devices through the Bluetooth or USB interface Color display...

Страница 17: ...he reagent strip and is more intelligent Auto update of memory Up to 1200 samples can be stored in memory You can query or via Bluetooth interface send test results to an external device When results...

Страница 18: ...sitive result 1 2 Intended Use The instrument can be used in combination with urine reagent strips for semi quantitative or qualitative detection of biochemical components in urine samples which can p...

Страница 19: ...It can t be used directly as evidence for diagnosing diseases 1 3 Contraindication None 1 4 Specifications 1 4 1 Performance of the instrument n Reagent Strips HC 2U HC 3U HC 4U HC 6U HC ACR HC 4K HC...

Страница 20: ...ing Mode Automatic single measuring mode n Display Color display screen which can display operational information of the instrument and test results etc n Memory Up to 1200 samples n Specific gravity...

Страница 21: ...29 mm L W H n Weight 190 g No battery n Power Supply Two AA alkaline batteries or 5 V 3 A The adapter Input 100 V 240 V 50 60Hz n Operating frequency 2 402 2 480GHz n Maximum transmitting power 4dBm...

Страница 22: ...V 3 A The adapter Input 100 V 240 V 50 60Hz n Power Consumption 13 5 W n Environment Friendly Use Period 10 years 1 5 Structure and Principle The instrument is basically composed of optical electronic...

Страница 23: ...ght source falls on the reagent pads on the strip The absorbance and reflectance vary with the color development of reagent pad i e The degree of color development is proportional to the concentration...

Страница 24: ...nsor system where the optical signals are transformed into electric signals Then the electric signals are transformed by I V convert then processed by CPU Finally test results are displayed on color d...

Страница 25: ...2 0 Printer figure 1 3 Figure 1 3 No Items Function 1 USB interface a For charging Charging voltage is 5 V DC b For transmitting data 2 Display screen Displays operational information test results etc...

Страница 26: ...cover Two AA alkaline batteries can be placed when turned on 5 Tray conveyor Used to place the strip tray 6 Strip tray Used to place the reagent strip 7 Printer cover Open this cover to load a roll of...

Страница 27: ...tch Long press for 3 seconds to start up After start up press this key into paper skip status When turned on long press for 3 seconds to shutdown 11 The printer reset key The printer will be re start...

Страница 28: ...tents for quantity and visible signs of damage according to the accompanying Packing List 3 Please notify your local distributor immediately if any damage or loss exists 2 2 Instructions for use NOTE...

Страница 29: ...present synthetic clothing carpets etc may cause damaging electrostatic discharges that may cause erroneous results b Do not use this instrument in close proximity to sources of strong electromagnetic...

Страница 30: ...peration of the device d Do not use this device in close proximity to sources of strong electromagnetic radiation e g unshielded intentional RF sources as these can interfere with the proper operation...

Страница 31: ...equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there...

Страница 32: ...d radio TV technician for help CAUTION To comply with the limits of the Class B digital device pursuant to Part 15 of the FCC Rules this device is to comply with Class B limits All peripherals must be...

Страница 33: ...ons of RE Directive 2014 53 EU A copy of the full DoC is attached 2 3 Installing the battery Caution 1 When installing the battery pay attention to the correct installation of the battery according to...

Страница 34: ...attery bin 3 Cover the battery cover 4 Press the start up key to start the instrument the instrument can start the work indicating that the battery installation is correct 2 4 Connect printer 1 Connec...

Страница 35: ...Thermal Print Paper Figure 2 2 1 Open the printer cover in the direction of upward as shown in figure 2 2 2 Use your fingers slight lift the printer rubber roller Printer cover Rubber roller Paper exi...

Страница 36: ...printing paper roll storage area then closed the printer cover Caution 1 Pay attention to the positive and negative of thermal printing paper Print characters display only in the front of the paper r...

Страница 37: ...3 2 check the print paper whether dry if the print paper is moisture then need to be changed in order to avoid the print paper jam 5 Please reinstall if appears print paper jam situation...

Страница 38: ...id exposure to urine samples 2 Use fresh urine samples Collect within one hour Samples must be measured within one hour otherwise keep them in refrigerator to avoid deterioration 3 Samples should be r...

Страница 39: ...ome test items will not get accurate test results 5 Sample volume should be enough to entirely soak all reagent pads on the strip You can t do the normal test if there are no enough samples 6 Directly...

Страница 40: ...urate test results 3 2 Reagent Strips Usage Precautions 1 Please use the special reagent strips Use only the URIT reagent strips and carefully read through the package inserts of the reagent strips 2...

Страница 41: ...ng so may affect test results 5 Please keep desiccant Do not discard the desiccant in the reagent strip bottle before using up all reagent strips Otherwise the reagent strip may go bad because absorb...

Страница 42: ...p consistent with the environmental conditions required by the urine strip used Caution The accuracy of the test results cannot be guaranteed if the instrument s operating environment fails to meet th...

Страница 43: ...can t guarantee the accuracy of the test results 2 Please do not wait for the battery to run out of charge before charging otherwise it will affect battery life Please through observe the battery ico...

Страница 44: ...lcome interface the system self test the tray conveyor extends from the front of the instrument The functions of each icon of the instrument are as follows Project Function 08 12 59 Display the system...

Страница 45: ...he power Keep pressing the Start key for 3 seconds to turn off the power The tray conveyor will retract inside the instrument and the display will be turned off Caution The instrument will automatical...

Страница 46: ...placed on a level stable and vibration free worktable 2 Do not touch the tray conveyor when it is moving 3 Verify that the use environment is compliant with the requirements when testing 4 Do not move...

Страница 47: ...agent strip Caution Do not centrifugal separation otherwise will cause haemocyte precipitation and make some items can t get accurate test results 4 2 3 Preparing reagent strips Please refer to the pa...

Страница 48: ...samples After you prepare the samples and reagent strips please soaking the reagent strips with samples and then do the tests The soaking time should use seconds as unit so you should read the follow...

Страница 49: ...to make sure all the reagent strips are soaked in the sample If the reagent block was not fully soaked some of the items could not get the accurate test results 2 The reagent strip should be soaked fo...

Страница 50: ...reagent block facing up 6 Insert the strip tray into the tray conveyor of the instrument and the instrument will test automatically Note that the handle end of the reagent strip should be on the outs...

Страница 51: ...e to the front of the instrument The results will be transmitted to external devices via the Bluetooth interface if connected 8 Take out the test tray 4 2 5 When the measurement is completed 1 The use...

Страница 52: ...results fall outside the range use another check strip to repeat the test If you continue to get out of range results the instrument may not be working properly Do not use the instrument Please conta...

Страница 53: ...he instrument can be normal use if not please check the check strip is all right or not 5 The range printed on the check strip container can only be used as the basis for judging instruments are in go...

Страница 54: ...d be guaranteed to be on the outside of the instrument If the reagent strip is in the wrong direction the instrument will prompt for the error 2 Be sure to place the reagent strip on the strip holder...

Страница 55: ...ected to the instrument 6 2 Test Function Click the second button on the right at the main interface to select the range for automatic identification of reagent strip type If testing 14TA reagent stri...

Страница 56: ...lied to screening some related diseases in the group according to the detection results of human urine It can t be used directly as evidence for diagnosing diseases 6 3 Introduction of Testing Result...

Страница 57: ...system will display the corresponding test items based on the type of reagent strip detected The test results are represented by a semi quantitative symbol See Appendix for a comparison table between...

Страница 58: ...via the Bluetooth interface if connected B If the instrument is successfully connected to the printer the test results will be printed C Connect the computer through USB and export the test results to...

Страница 59: ...click Setting key under the standby screen into setting menu which has following functions 1 QR code Scan the QR code and connect to the mobile APP 2 Unit You can choose the unit is Char SI Conv Char...

Страница 60: ...date firmware After clicking this button a dialog box appears on the screen which displays connect USB cable to device and PC Copy firmware Press Yes When firmware is copied to the instrument press Ye...

Страница 61: ...ument and app D Click synchronize data in app data center interface to synchronize data 2 Mobile phone with iOS system A Turn on Bluetooth When the iPhone connects to the instrument for the first time...

Страница 62: ...5 7 6 7 Mobile APP introduction The mobile app is divided into three modules Device Management Health Assessment and Data Center As shown in figure 6 2 Figure 6 2...

Страница 63: ...gement interface 2 Health Assessment The latest comprehensive health assessment is displayed in the Health Assessment interface 3 Data Center In the data center you can synchronize all the test result...

Страница 64: ...nd protective gloves according to local regulations 7 2 General cleaning Keep the instrument clean and dust free If cleaning is needed clean the instrument surface with clean paper or gauze dipped wit...

Страница 65: ...are completed 1 Turn on the power and wait for self check of the instrument 2 Please remove the strip tray from the tray conveyor 3 Clean the strip tray Clean the strip tray by using mild detergent a...

Страница 66: ...utmost care and avoid moisture sunlight and collision Transportation Condition 20 55 RH 95 75 kPa 106 kPa 8 2 Storage The packed instrument should be stored in a ventilated room Do not store the inst...

Страница 67: ...alers or after sales service unit immediately Caution 1 If an fault message appears clear the error and retest the sample using a new reagent strip Correct results may not be yielded if the soaking ti...

Страница 68: ...is falling off Restart the instrument If still unable to resolve please contact the after sales service unit T 2 Strip Motor Error Check if the strip tray is installed correctly T 3 Strip Type Error P...

Страница 69: ...6 4...

Страница 70: ...6 5 Appendix The instrument test results and the density contrast compare table...

Отзывы: