VERSAMET 4 INVERTED METALLURGICAL MICROSCOPE
UNITRON
73 Mall Drive, Commack, NY 11725 • 631-543-2000 • www.unitronusa.com
20
TROUBLESHOOTING
(continued)
MECHANICAL PART
PROBLEM
CAUSE
SOLUTION
The coarse adjustment knob is
too difficult to rotate
The tension adjustment ring is tightened too
much
Loosen it
The image goes out of focus
during observation or the stage
drops by itself
The tension adjustment collar is too loose
Tighten it
ELECTRICAL SYSTEM
PROBLEM
CAUSE
SOLUTION
The lamp
doesn’t light
No power to the lamp
Check the power cord is connected correctly
The lamp is not installed correctly
Install it correctly
The lamp burns out
Replace the lamp
The light intensity is not
enough
The light intensity control knob is not set
properly
Adjust the light intensity control knob
The wrong lamp is used
Replace the lamp with the correct one
The lamp keeps burning out
The wrong lamp is used
Replace the lamp with the correct one
The lamp flickers or the
brightness is not stable
The lamp will burn out soon
Replace the lamp
The power supply
doesn’t connect well
Connect the power supply correctly
MISCELLANEOUS
When using a camera and
viewing on a monitor, a
reflection from a window or
indoor lighting is seen in the
image
The stray light entered through the eyepieces
or viewfinder is reflected
Cover both eyepieces to prevent stray light
from entering