The Versamet 4 is outfitted with a binocular viewing
head with one photo port for HDMI/digital imaging.
You must select the appropriate light path for
observing specimens.
The light path is set to 100% to the binocular
eyepieces as the default setting at our facilities where
the light path selection slider
①
is set to the “OUT”
position (pushed all the way to the right).
Slide the light path selection slider to the left to the
“IN” position. This will send 0% of the light to the
binocular viewing head and 100% to the top photo
port for HDMI/digital imaging and documentation.
Adjusting the Focus
(Fig. 13-14)
To ensure that you obtain sharp images with both
eyes, (since eyes vary, especially for those wearing
glasses) any eyesight variation can be corrected in
the following manner.
Set both diopter collars
①
to “0” by aligning the line
on the eyepiece
②
with the line on the diopter collar.
Using your left eye only and the lowest magnification
objective (5x or 10X), focus your specimen by
adjusting the coarse adjustment knob
③
.
When the image is in view, refine the image to its
sharpest focus by turning the fine adjustment knob
④
.
Rotate the diopter collar to obtain the sharpest focus.
To obtain the same sharp image using your right eye,
do not touch the coarse or fine adjustments. Instead,
rotate the right diopter collar until the sharpest image
appears. Repeat several times to check.
IMPORTANT:
DO NOT counter rotate the focusing
knobs as this will cause severe problems and
damage to the focusing system.