VERSAMET 4 INVERTED METALLURGICAL MICROSCOPE
UNITRON
73 Mall Drive, Commack, NY 11725
• 631-543-2000 • www.unitronusa.com
18
OPERATION
(continued)
LAMP REPLACEMENT -- CAUTION
:
the glass housing of the lamp may be
extremely hot.
DO NOT
attempt to
change the lamp before it is completely
cooled or without wearing adequate skin
protection.
Replacement Lamp:
Catalog # 150-30-100
(12volt 100 watt halogen)
Using the DIC Components
(Fig. 25)
(continued)
The fine adjustment knob
③
can control the background
interference color change from gray to mauve.
NOTE:
Adjusting the aperture diaphragm at the same
time will provide the best contrast.
Since the differential interference contrast is very
sensitive, there should not be any dirt or dust on the
surface of the sample.
Birefringent materials cannot reach the effect of
differential interference contrast in microscopical
examination.
Replacing the Lamp
(Fig. 26-27)
The halogen lamp was installed by our factory technicians
prior to shipment. Should you need to change the lamp,
following the instructions below.
NOTE:
When handling the halogen lamp, handle it
only using lint-free gloves, cloth or tissue. Touching
the lamp with your fingers or other type of material
can damage the bulb.
Turn the power switch to OFF and UNPLUG the
microscope power cord; allow the lamp to cool.
Using the hex screwdriver supplied with your
microscope, loosen the lock screw
①
and remove the
black illuminator housing cover by sliding it upwards as
shown.
Using a lint-free glove, cloth or tissue, gently grasp the
lamp
②
and push the lamp release lever
③
back to
release the lamp; pull the lamp out.
Using a lint-free glove, cloth or tissue, gently grasp the
replacement lamp, push the lamp release lever back
and insert the lamp pins
④
into the jack
⑤
closest to
the hinge of the lamp release lever, then release the
lever; the lamp should be horizontal after installation.
Replace the illuminator housing cover and tighten the
lock screw.
Fig. 26
①
②
⑤
④
Fig. 27
③