17
Moxi
2
battery door replacement
Removal
1. Use the small silver tool to push the pin out.
2. Remove the battery door.
Installation
1. Line up the battery door.
2. Insert the pin. Use the back of the small silver tool
to push the pin in.
How to measure performance in a coupler
RIC devices should be measured with a HA1 coupler
since it best reflects the acoustical properties in the ear
canal.
The xReceiver kit includes a metal adapter plate that
fits on a HA1 coupler and allows for a good connection
of the xReceiver unit to the coupler.
Select
Activate Test Box Measurement Mode
in
Unitron TrueFit software from the menu on the top
menu bar. Clicking on
Activate Test Box Measurement
Mode
disables the adaptive features and activates
the omnidirectional microphone mode to allow for
correct measurements. Click on
Deactivate Test Box
Measurement Mode
to reestablish the previous
settings.
Attach the standard xReceiver to the metal adapter plate.