TROUBLESHOOTING 

Under certain conditions, performance of this unit may be adversely affected by factors other than 
defects. If a problem occurs, please review the following list and take remedial action as needed. If you 
cannot solve the problem after checking the entire list, please contact your local dealer for assistance. 

 

Trouble 

Cause 

Remedy 

Double images 

Interpupillary distance is not correct 

Readjust it  

Diopter adjustment is not correct 

Readjust it  

 
 
Dirt appears in the view field 

Dirt on the specimen 

Clean specimen 

Dirt on the surfaces of eyepieces  

Clean eyepieces 

Dirt on surface of objective 

Clean objective 

Dirt on stage 

Clean stage 

Image is not clear 

Dirt on the surface of objectives 

Clean objectives 

Image is not clear while focusing  

Diopter adjustment is not correct 

Readjust diopter 

Focusing is not correct 

Readjust it 

The focusing knob is not smooth 

The focusing knob is too tight 

Loosen it to a suitable position 

The image is obscure because of the 
head slipping down during 
observation 

The focusing knob is too loose 

Adjust the tension on the focus 
knobs 

 
 
 
Lamp does not work 

Bulb may be loose in socket 

  Remove bulb and reinsert 

securely 

Bulb has burned out 

Replace the bulb 

Fuse has burned out 

Replace the fuse 

Dirt on the stage obstructed the path 
of light 

Clean the stage 

 
Lamp burns out frequently 

Voltage from power supply is too 
high 

Use the brightness adjustment 
knob to lower brightness 

The bulb is not standard 

Use the recommended bulb  

Fuse burns out frequently 

Voltage from power supply is too 
high 

Use the brightness adjustment 
knob to lower brightness 

Light flickers 

Bulb needs replacement 

Replace bulb 

Bulb may be loose in socket 

  Remove bulb and reinsert 

securely 

 

 

Содержание Gemscope 3075-GS

Страница 1: ...73 Mall Drive Commack NY 11725 631 543 2000 P 631 589 6975 F www unitronusa com info unitronusa com Gemscope MANUAL ...

Страница 2: ...itronusa com 2 CONTENTS SAFETY NOTES 3 REPAIR CARE AND MAINTENANCE 4 INTRODUCTION 5 UNPACKING AND COMPONENTS 5 COMPONENTS DIAGRAM 5 ASSEMBLY 6 8 ADJUSTMENT AND OPERATION 9 11 REPLACING THE ILLUMINATOR LAMPS 12 13 TROUBLESHOOTING 14 MAINTENANCE 15 SERVICE 15 WARRANTY 15 ...

Страница 3: ...nge the lamp before it is completely cooled or without wearing adequate skin protection 5 All electrical connectors power cord should be inserted into an electrical surge protector to prevent damage due to voltage fluctuations 6 FUSE REPLACEMENT For safety when replacing the fuse ONLY replace with the same size type and rating of original fuse be sure the main switch is in the off position disconn...

Страница 4: ...ularly Accumulated dirt on metal surfaces should be cleaned with a damp cloth More persistent dirt should be removed using a mild soap solution Do not use organic solvents for cleansing The outer surface of the optics should be inspected and cleaned periodically using an air bulb If dirt remains on the optical surface use a soft lint free cloth or cotton swab dampened with a lens cleaning solution...

Страница 5: ...emscope is of the highest quality prior to shipment UNPACKING AND COMPONENTS Your Gemscope arrived packed in a molded shipping carton Do not discard the carton the shipping carton should be retained for reshipment of your Gemscope if needed Avoid placing the Gemscope in dusty surroundings or in high temperature or humid areas as mold and mildew can form Carefully remove the Gemscope from the shipp...

Страница 6: ...m below shows how to assemble the various components The numbers indicate the order of assembly When assembling the Gemscope make sure that all parts are free of dust and dirt and avoid scratching any parts or touching glass surfaces Stereo Head Lock Screw Power Cord Eyepieces Fluorescent Illuminator Gem Clamp ...

Страница 7: ...d insert the stereo head into the bracket of the stand Tighten the lock screw to secure the stereo head Installing the Eyepieces Fig 1 Remove the eyepiece dust caps and insert the eyepieces until they stop as shown in Fig 1 Installing the Stoneholder Fig 2 The stage has two stoneholder mounts on either side of the stage where you can attach the stoneholder Fig 1 Fig 2 ...

Страница 8: ...ace Hand tighten the connector If possible it is recommended to further tighten the connector by using rubber tipped pliers This will ensure that the gooseneck will not loosen and rotate as the overhead fluorescent light is being used Installing the Optional LED Light Fig 3A Remove the plastic cap covering the opening to the LED socket Insert the LED light guide into the socket and seat it as deep...

Страница 9: ... left and right focus adjustment knobs with both hands hold the left knob to prevent it from turning and rotate the right knob clockwise to increase tighten or counterclockwise to decrease loosen the focus knob tension After tension adjustment has been completed always rotate both adjustment knobs in the same direction Adjusting Interpupillary Distance Fig 6 Different users have different interpup...

Страница 10: ...when users change because different users have differing diopters Rotate the magnification selector knob to 3x and rotate the focusing knob to focus on the iris diaphragm Rotate the magnification selector knob to the 1x magnification position and adjust the diopter ring on left eyepiece to focus the specimen repeat for the right eyepiece Adjusting the Bottom Illumination Fig 8 Plug the GEM base in...

Страница 11: ...rotating the baffle control knob counterclockwise Inclusions should stand out as dark objects against a bright background Darkfield Illumination Close the baffle by rotating the baffle control knob clockwise The stone is illuminated with diffused light from the sides View the stone against the baffle s black non reflective background Inclusions stand out as bright objects Adjusting the Top Fluores...

Страница 12: ... housing is a trap door Open the trap door by loosening the screw see Fig 11 The screw can be loosened by hand or by using the small flathead screwdriver included in the original shipment of the Gem stand The door will open and the lamp can be removed IMPORTANT When replacing the lamp be careful not to touch the new lamp with your fingers as the oils and dirt on your fingers can affect the lamp s ...

Страница 13: ...r plate using the Phillips head screwdriver that was included with your Gem stand The fluorescent lamp is connected to a socket in the housing by a four pin connector The socket is held in place with a screw Remove the screw from the lamp socket Carefully pull the fluorescent lamp out from the socket by the base of the lamp to avoid accidental breakage Insert the new lamp into the socket and repla...

Страница 14: ...jectives Image is not clear while focusing Diopter adjustment is not correct Readjust diopter Focusing is not correct Readjust it The focusing knob is not smooth The focusing knob is too tight Loosen it to a suitable position The image is obscure because of the head slipping down during observation The focusing knob is too loose Adjust the tension on the focus knobs Lamp does not work Bulb may be ...

Страница 15: ...r This warranty does not cover damage caused in transit misuse neglect abuse or damage resulting from improper servicing or modification by other then UNITRON approved service personnel This warranty does not cover any routine maintenance work or any other work which is reasonably expected to be performed by the purchaser Normal wear is excluded from this warranty No responsibility is assumed for ...

Отзывы: