Range
Resolution
Accuracy (45-1kHz)
0,50...60,00 mA
0,01 mA
± 1,5 % rdg. +5 D
60,00...99,99 mA
0,01 mA
± 1,5 % rdg. +50 D
100...600,0 mA
0,1 mA
± 1,5 % rdg. +5 D
0,050...3,000 A
0,001 A
± 1,5 % rdg. +5 D
3,000...9,99 A
0,001 A
± 1,5 % rdg. +50 D
10,00...30,00 A
0,01 A
± 1,5 % rdg. +5 D
30,00...50,00 A
0,01 A
± 1,5 % rdg. +5 D
Overvoltageprotection: 600 A AC
Harmonics in V Measurement Range (Percentage)
specified from 80 V rms AC 50/60Hz, max. Peak Voltage 850 Vss
Range
Resolution
Accuracy (50/60Hz)
1...10
0,1 %
± 1 % rdg. ± 1%
11...20
0,1 %
± 5 % rdg. ± 1%
21...50
0,1 %
± 15 % rdg. ± 1%
51...99
0,1 %
± 35 % rdg. ± 1%
Harmonics in V Measurement Range (Absolute Value)
specified from 80 V rms AC 50/60Hz, max. Peak Voltage 850 Vss
Range
Resolution
Accuracy(50/60Hz)
1...10
0,1 V
± 1 % rdg. ± 7 D
11...20
0,1 V
± 5 % rdg. ± 7 D
21...50
0,1 V
± 15 % rdg. ± 7 D
51...99
0,1 V
± 35 % rdg. ± 7 D
Harmonics in mA, A Measurement Range (Percentage)
specified from 100 mA rms 50/60Hz (mA Range) or from 1A rms 50/60Hz (A Range),
max. Peak Voltage 850 mAss (mA Range) or 70 Ass (A Range)
Range
Resolution
Accuracy (50/60Hz)
1...10
0,1 %
± 1 % rdg. ± 1%
11...20
0,1 %
± 5 % rdg. ± 1%
21...50
0,1 %
± 15 % rdg. ± 1%
51...99
0,1 %
± 35 % rdg. ± 1%
15
PTDB93570002.qxd:Layout 1 30.05.2008 15:05 Uhr Seite
Содержание CHB 5
Страница 1: ...CHB 5 Leckstrommesszange 93570 Bedienungsanleitung Instruction Manual...
Страница 2: ......