Model No. : UT-101 F
PAGE : 7
This Mode is used to Calibration of the Instrument.
CALIBRATION MODE PARAMETER
CAL
3.
CALIBRATION MODE
Parameter
Description
Range
Default Value
Password for Parameter Setting
Mode
0000
0000
Scroll the digit by
key, increase by
key.
Calibration Span.
CALs
----
----
Values between
0000
to
9999
According to Input type.
Password for Calibration Parameter
Calibration Zero.
CALZ
----
----
According to Input type.
Apply low signal (from input table) input
and press
key.
Apply High signal (from input table) input
and press
key.
Parameter
Description
Range
Default Value
ofst
0000
Values from
to
000.0
999.9
000.0
Offset to Process Value.
Scroll the digit by
key, increase by
key.
This value Increase OR Decrease the Display PV
Output Low
oP-L
0000
Increase digit by
key, decrease by
key
Process Value at 4 mA signal output.
Values from
to
000.0
999.9
000.0
Output High
oP-H
1000
Increase digit by
key, decrease by
key
Process Value at 20 mA Signal output.
Values from
to
100.0
999.9
000.0
Hys1
0002
Increase digit by
key, decrease by
key
Values from
to
000.2
999.9
000.0
Hysteresis # 2
for Thermocouple(T/C)
set 0.0mV for Zero Cali.
for Thermocouple(T/C)
set 42.0mV for E,J,K-T/C for
set 18.0mV for R,S,T-T/C for
Span Cali.
Span Cali.