20
20
Problem
Sensor has a good zero signal but has a very low response
to N2O.
Possible cause
Air bubble in the narrow parts of the sensor.
Solution
Shake the sensor as you would shake an old school
mercury thermometer. See video guide here:
unisense.com/
video-guides/
Problem
A slow response
Possible cause
Insoluble compounds deposited at the sensor tip
Solution
Rinse with 96% ethanol, rinse with 0.01M HCl and rinse with
water
Problem
An unstable signal or the signal fluctates if the setup is
touched or equipmen is introduced in the medium you are
measuring in
Possible cause
Electrical disturbance of the sensor through the tip
membrane
Solution
Ground the set-up using the blue grounding cable supplied
with the picoammeter. Connect the reference plug on
the picoammeter (blue plug) with the medium you are
measuring in.
Problem
High and drifting signal
Possible cause
The sensor tip is broken
Solution
Replace the microsensor
TROUBLESHOOTING
Содержание N2O-50
Страница 1: ...NITROUS OXIDE SENSOR USER MANUAL...
Страница 2: ...NITROUS OXIDE SENSOR MANUAL Copyright 2024 Unisense A S Version January 2024...
Страница 3: ...NITROUS OXIDE SENSOR MANUAL UNISENSE A S...
Страница 23: ...www unisense com sales unisense com...