16.D/A CONVERTER(OPTION)
104
How to output adjust
Adjust the output with the dummy trimmer that appears on the adjustment screen after
selecting each fixed output in the D/A output mode setting and pressing the [OK] key.
While monitoring the output value, finely adjust the output with the dummy trimmer UP/
DOWN keys, and press the [OK] key to confirm, so that the trimmer position is entered.
*Pressing the dummy trimmer UP/DOWN keys continuously performs continuous
operation.
*The trimmer adjusting ranges are as follows: voltage output; about
±
2.0V, and current
output; about
±
1.6mA.
*The reset key resets the dummy trimmer position to the trimmer center point (0%).
Reset key
Dummy trimmer
position indicator
Dummy trimmer adjustment
UP/DOWN keys