Chapter
2
Precautions about connection to the signal input/output terminal block are given below.
The precautions described here are important for safety.
Please properly understand the descriptions before connection.
WARNING
- Do not connect a commercial power source directly to the signal input/output
terminals.
- Connect to the signal input/output terminals with no power applied because it may
cause an electric shock.
- For connection to the signal input/output terminals, check the signal names and pin
assignment numbers, and then carry out wiring properly.
- After wiring, be sure to mount the attached terminal block cover. Otherwise, it may
cause an electric shock.
- Before applying power, carefully check the wiring, etc.
- Do not touch any signal input/output terminal while applying power.
Doing so may cause electric shocks or malfunctions
CAUTION
- Tighten the terminal screws at the specified torque.
If they are loose, shorts, fire or malfunctions may occur. (When D/A converter, RS-
232C, or RS-485 option is installed)
Tightening torque: 0.2 to 0.6 N•m
- Use shielded cables.
2
CONNECTION
Содержание F331
Страница 1: ...OPERATION MANUAL 26MAY2014REV 1 02 DIGITAL INDICATOR F331 ...
Страница 79: ...70 8 SPECIFICATIONS 70 SPECIFICATIONS Chapter 8 8 2 Dimensions Unit mm ...
Страница 80: ...71 8 SPECIFICATIONS 71 SPECIFICATIONS Chapter 8 When the BCD output option is equipped Unit mm ...
Страница 81: ...72 8 SPECIFICATIONS 72 SPECIFICATIONS Chapter 8 When the D A converter option is equipped Unit mm ...
Страница 82: ...73 8 SPECIFICATIONS 73 SPECIFICATIONS Chapter 8 When the RS 485 option is equipped Unit mm ...
Страница 83: ...74 8 SPECIFICATIONS 74 SPECIFICATIONS Chapter 8 When the RS 232C option is equipped Unit mm ...