background image

PowerSpin C8624 Centrifuge 

Parts and Nomenclature 

 

 
 
 
          
 
           Lid 
 
 
 
 
 
 
 
 
 
 
 
          Rotor 
 
 
 
 
Emergency 
lid-opening  
(left side) 
 
Control 
panel 
 
 
 

 
 

      

 

                                                      Start, Stop, Open buttons 
                                                   
        Digital Timer & Speed Control  
 

 
 
 
 
Safety 
lock 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
On/off 
switch 
(on back 
side) 
 
 
 
Power 
indicator

 
 
                                                        Model C8624  

 11

Содержание PowerSpin C8624

Страница 1: ...PowerSpin C8624 24 Place Centrifuge OPERATION MANUAL 182 Ridge Road Suite E Dayton NJ 08810 USA 800 588 9776 Fax 732 274 1151 Web www unicosci com...

Страница 2: ...tal display simultaneously Electronic 60 minutes timer with beeping signal at end of run Automatic time and speed reset to the last time speed and time setting for repeated job Memorized time and spee...

Страница 3: ...shion foam in the rotor chamber Do not operate the centrifuge before the foam is removed 1 Take the centrifuge from the Styrofoam and place it on a sturdy flat surface Do not turn on the centrifuge 2...

Страница 4: ...conds To increase time setting press and hold up arrow otherwise down arrow button Speed Setting and Speed Setting Change Speed setting Use the up or down arrow to preset the running speed The speed r...

Страница 5: ...total to date number of cycles will be displayed Note Cycles of usage is not total spinning time It represents the total times the centrifuge has spun Operation 1 Plug the centrifuge into a power sou...

Страница 6: ...is not necessary A surface cleaner such as Formula 409 can be used to clean the outside of the instrument The rotor chamber and cover can be disinfected with a mild solution of bleach or other disinfe...

Страница 7: ...switch in the back of unit is in the on position e Check that the power indicator light with the switch is lit A failure of the power light to illuminate could indicate a need for a new fuse when the...

Страница 8: ...which is 250V 4A e Put the snap in fuse holder back 5 Intermittent motor operation or no operation Contact your representative for possible motor replacement Specifications Speed 1 000 3 400 RPM Spee...

Страница 9: ...ed distributor or dealer within five 5 days after the occurrence of said defect No other warranty of any kind is made expresses or implied The warranty described above shall be the sole and exclusive...

Страница 10: ...o response no response sound 1 Loose connection or connection reversed 2 Touch panel defective 1 Exercise the connector J2 on main 2 Replace touch panel 6 Display show Running but motor is not running...

Страница 11: ...4 Centrifuge Parts and Nomenclature Lid Rotor Emergency lid opening left side Control panel Start Stop Open buttons Digital Timer Speed Control Safety lock On off switch on back side Power indicator M...