Button / Connector   

Description 

/  

Press to toggle between Celsius and Fahrenheit 

MAX/MIN 

Press to toggle between MAX and MIN options.    MAX and 
MIN values are displayed on the secondary display. 

Press to toggle between turning on the laser and display 
backlight. 

The sequence is: 

,

,

 none.

 

 

Locating a Hot or Cold Spot 

To find a hot or cold spot, aim the Thermometer outside the target area.    Then, slowly scan across the area with an up and down motion until you located the hot or cold spot. 
See Figure 5.

 

 

 

 
 

Distance and Spot Size 

As the distance (D) from the target being measured increases, the spot size (S) of the area measured by the unit becomes larger.    The spot size indicates 90% encircled 
energy.    The maximum D:S is obtained when the Thermometer is 1000mm (100 in) form the target resulting in a spot size of 20mm (2 in).   

 

 

 

Field of View 

Make sure that the target is larger than the spot size.    The smaller the target, the closer you should be to it. Suggested distance less than 75% of the theory value.

 

 

 

 

 

 

Emissivity 

Emissivity describes the energy-emitting characteristics of materials.    Most organic materials and painted or oxidized surfaces have an emissivity of about 0.95. 
 
If possible, to compensate for inaccurate readings that may result from measuring shiny metal surfaces, cover the surface to be measured with masking tape or flat black paint 
(<150   / 302 ) and use the high emissivity setting.    Allow time for the tape or paint to reach the same temperatures as the surface beneath it.    Measure the temperature of 
the tape or painted surface. 
 
 

Maintenance  

 
Changing the Battery 

To install or change the 9V battery, open the battery compartment the battery. 
 

Cleaning the Lens 

Blow off loose particles using clean compressed air.    Carefully wipe the surface with a moist cotton swab.    The swab may be moistened with water. 
 

Cleaning the Housing 

Use soap and water on a damp sponge or soft cloth.    To avoid damage to the Thermometer, do not soak the unit into water. 
 

 Caution 

To avoid damaging the Thermometer, do NOT submerge it in water. 

 

Troubleshooting 

Symptom Problem 

Action 

OL (on display) 

Target temperature is over range 

Select target with specifications 

-OL (on display) 

Target temperature is under range 

Select target with specifications 

 

Low Battery 

Replace Battery 

Blank Display 

Possible dead battery 

Check and / or replace battery 

Laser does not work 

z

 

Low or dead battery 

z

 

Ambient temperature above 40   (104 ) 

z

 Replace 

battery 

z

  Use in area with lower 

ambient temperature. 

 

CE Certification 

The Thermometer conforms to the following standards:

 