UTG1000A Series Function Generator User Manual
18
3.3.4
Square Wave Setting
Press Menu→Waveform→Type→Squarewave→Parameter in turn (press Type softkey to select only
when Type label is not highlighted). If parameter needs to be set, press corresponding softkey to enter
required numerical value and select the unit.
Note: This parameter can be set by multifunctional knob and direction buttons.
3.3.5
Pulse Wave Setting
Default duty cycle of pulse wave is 50% and rising/falling edge time is 1us. Steps for setting square
wave with 2ms period, 1.5Vpp amplitude, 0V DC offset and 25% duty cycle (limited by the minimum
pulse width specification 80ns), 200us rising time and 200us falling time are seen as following:
Press Menu→Waveform→Type→PulseWave→Parameter in turn, then press Frequency softkey to
switch to Period. Enter required number value and select the unit. When entering duty cycle value,
there is a quick label at the bottom of display, and select 25%.
If need to set falling edge time, press Parameter softkey or rotate multifunctional knob to the right to
enter sub label, then press Falling Edge softkey to enter required number and select unit.