UT3500 series
User Manual
37
6. After the data statistics are automatically started, you can use the function key to stop
the current statistics at any time; you can also save the data to the U disk at any time;
you can also stop and clear the cached data, as shown in Figure 6-2-3.
Figure 6-2-3 Statistical Operation Interface Display
6.2.3 Save Data to U Disk
After the data stat is enabled, use the same operation method as the above [LOG]. The
data cache can be cleared through the function keys at the bottom of the screen, and the
data can also be saved to a U disk at any time.
The data statistics function can only be set after opening [LOG/STAT] on the
<SYSTEM CONFIG> page. After the data statistics function is turned on, the instrument
needs to perform complex calculations on multiple parameters, and the measurement
speed will drop slightly. Once the data recording is started, the test page will be locked and
cannot be switched to other pages. In the external trigger state, before switching to other
pages, data logging must also be closed. If switching from other pages to the <TEST>
page, data logging will automatically start.
7. Comparator Sorting
This chapter mainly introduces how to use the beep and the comparator setting to realize
the production line judgment and prompt whether the product meets the specification
requirements. This is an important process for the realization of automatic assembly line
test screening.