UT3200 Series User Manual
22
Open comparator steps:
1.
Press Setup enter Function page
2.
Move the cursor key to the high and lower limit temperature of specific channel. Choose
start to open comparator, screen will display COMP
3.
Move the cursor key to turn on/off audible alarm, when test value exceed comparator
range, audible alarm will alert you
4.
User can also set the corresponding sampling rate or temperature unit according to their
needs. Specific setting steps can refer to the section on Functional Settings
5.2 Channel setting
Open comparator to set the high and lower temperature limit. Refer to Figure 5-2-1
Figure 5-2-1
Steps:
1.
Through Function page or Test page to enter Channel Setting
2.
Move the cursor key to the upper and lower limit temperature of specific channel, use the
numeric key enter the upper and lower limit value then press OK to confirm. See Figure 5-
2-1
3.
User can use Function button to make set according to their needs
Function
Description
Rest
The upper limit of the current channel
return to the factory setting
One key setting
Set the upper limit of other channel as same
as current channel
4.
The steps for setting up other channels are the same as above
5.
Tc model setting is for other temperature connection line replacement for each channel
sensor. User can set refer to the following table information to make set
Function
Description
TC-K
K thermocouple
TC-T
T thermocouple
TC-J
J thermocouple