6
5.
Installation
5-1.Location of Installation
Place it horizontally near a vacuum pump.
Circular pumps are of pressurized type. Sucking failure or other failures may occur if the
connecting hose in IN side is too longer or if the interval to the pump is too wide. The
following locations are not proper for installation.
①
Vibrating places
②
The places with flammable vapors or particular gas.
③
The places under high temperature/humid
④
The places possibly might suffer from damage by heat/water
5-2.Connection to a vacuum pump
GCD series’ models only are adoptable for our oil rotary pump.
Additional set up is required when we adopt the other models. Please contact us for
further in detail.
Follow instructions at the previous page, P5 drawings and Section 6 for connecting to a
vacuum pump.
5-3. Oil and water separating element
If the element is soaked with water beforehand for 1 to 2 minutes, it helps form the water
separation membrane and improves its function at the operation outset. Always soak with
water before use.
To remove, refer to “10. Replacement of Elements.”
6.
Preparation for operation and supply
6-1
.
Place this body and oil rotary vacuum pump at the height of the surface of oil or lower
than it, within 1m. The oil in the body may reflux (back run) towards the pump side, if
you set the body higher than the pump.
6-2
.
Connect the body to the oil rotary vacuum pump as described in Section 4.
1)
Mount brass hose nipple of accessory in oil inlet and outlet port of GCD series models.
2)
Connect between inlet port
①
and oil drain port of vacuum pump.
Fix it with hose band firmly.
3)
Connect accessory tube between oil drain port
③
and oil supply port of vacuum pump.
Attach it together with hose band tightly.
6-3
.
For oil supplement, install the tube into the oil refill port of the Inlet side 3-way
valve.
6-4
.
Prepare oil for supply beforehand. (Refer to UFW-003 for the oil quantity of IL and
vacuum pump)
Length of connecting hose
Within 1.5m
Vertical interval to the pump
The same level or within a range of lower level of 1m