Safety Precautions

 

 

In order to safely use the G-TRAN Series sensor unit (below, "this unit"), please read the instruction 

manual and the safety precautions below. 

 

Request repairs for this unit from the dealer where purchased, from ULVAC Inc., or 

from the URL listed in this instruction manual. 

Repair 

 

If this unit is accidentally damaged, immediately turn off the power supply. There is 

a risk of fire and electric shock from use in this state. 

Power off 

 

If this unit produces unusual heat, smoke, or a strange smell, immediately turn off the 

power supply. There is a risk of fire from use in this state. 

Power off 

 

Before turning on the power supply, check that the operating voltage to the SH2 

multi-ionization gauge and the supply voltage are the same. If the wrong power supply 
is mistakenly connected, there is a risk of damage to the devices connected to the SH2 
multi-ionization gauge and a risk of fire. The overvoltage category is category I. 

Power supply voltage 

 

Do not connect this unit in a location where the gauge head's internal pressure 

exceeds atmospheric pressure. If the gauge head's internal pressure exceeds 
atmospheric pressure, there is a risk of damage to the gauge head and it may pose a 
hazard to its surroundings, including to people, by flying off its connector. When 
exceeding atmospheric pressure, install an isolation valve to ensure that the gauge 
head's internal pressure does not exceed atmospheric pressure. 

Excessive vacuum chamber pressure 

 

Always connect this unit electrically to a grounded vacuum chamber. Use a 

conductive metal clamp on the NW flange. If this unit is not correctly connected, it will 
not only show the wrong pressure, there is a risk of damage to this unit and the 
equipment connected to this unit and a risk of fire. 

Protective grounding 

When isolated with the vacuum chamber, ground the NW flange metal clamp with a 

class D (class 3) ground. 

 

The pressure value and accuracy for this unit differ by the type of gas measured. The 

accuracy for this unit is based on measuring nitrogen gas. Be careful of 
overpressurization when using gases other than nitrogen. 

Gas type dependency 

 

Do not disassemble this unit. There is a risk of fire and electric shock. If 

disassembled, sensor operation cannot be guaranteed. 

Disassembly prohibited 

 

Do not modify this unit. There is a risk of fire and electric shock. If modified, sensor 

operation cannot be guaranteed. 

Modification prohibited 

 

Before turning on the power supply, check that this unit operating voltage and the 

supply voltage polarity are the same. If the wrong power supply is connected, there is a 
risk of damage to this unit and to devices connected to this unit and a risk of fire.

 

Check power supply voltage 

 

Always connect this unit to a power supply and evaluation unit that conforms to 

standards (EN61010 SELV-E) for safety extra low voltage protected by grounding. 
Install a fuse in the connection. 

Operating power supply 

 

After connecting this unit and the SH2 with the unit cable, turn on the power supply to 
the SH2. If the power is supplied before connecting them, there is a risk of damage to 
this unit and the SH2 and a risk of fire. There is also a risk of breaking the filament in 
the gauge head.

 

Power on 

Содержание SPU

Страница 1: ...Measuring Unit SPU Instruction Manual Components Division ULVAC Inc http www ulvac co jp This manual is for this unit units of the following serial numbers S N 00001 and higher Read this manual befor...

Страница 2: ......

Страница 3: ...ty Symbols Safety symbols are used throughout this instruction manual to call the operator s attention to safety The terminology used in safety symbols is classified below Indicate status of urgency o...

Страница 4: ...ly to a grounded vacuum chamber Use a conductive metal clamp on the NW flange If this unit is not correctly connected it will not only show the wrong pressure there is a risk of damage to this unit an...

Страница 5: ...zation of moisture or organic matter pressure accuracy and response speed will be affected Review installation locations and methods so that fouling does not occur Moisture and fouling The ambient tem...

Страница 6: ...ry There is a risk of damage if this unit is transported without packaging or transported attached to the equipment Transportation packaging When disposing of this unit please follow all local laws an...

Страница 7: ...3 2 LEAK TEST 6 4 HANDLING PRECAUTIONS 7 4 1 PRECAUTIONS RELATED TO THE USAGE ENVIRONMENT 7 4 2 PRECAUTIONS RELATED TO USAGE 7 4 3 ATTACHING THIS UNIT 7 4 3 1 This unit attachment environment 7 4 3 2...

Страница 8: ...e Measurement value output Output by serial communication LED display POWER ERROR LED Gauge head material Filament Pt Other CuZn Ni plating Kovar glass Ni SnSb Gauge head withstand pressure 2x10 5 Pa...

Страница 9: ...re 2 1 SPU controller section explanatory drawing Name symbol Function I O connector The I O connector for the power supply and signals RJ 45 jack connector Gauge head clamp The clamp to hold and lock...

Страница 10: ...igure 2 2 SPU LED explanatory drawing Name symbol Function POWER ERROR LED On Operating normally Flashing Error such as a filament break this unit is malfunctioning Off Power off this unit or the SH2...

Страница 11: ...ctor pin assignment diagram Terminal number This unit Signal direction Connection destination 1 Power supply 5 VDC SH2 Multi ionization gauge 2 GND 3 Power supply 5 VDC 4 Communication terminal 5 Comm...

Страница 12: ...e filament in the gauge head Power on Perform a leak test when the gauge head is replaced and attached There is a possibility of leaking from impacts to the pin section Leak test 3 1 Attaching the gau...

Страница 13: ...igure below the gauge head may leak from the constrictive clamp Insufficiently inserted Leak occurs Fully inserted 3 2 Leak test When the gauge head is attached we recommend performing a leak test in...

Страница 14: ...um chamber and protect this unit so it is not exposed to these gases as much as possible by using this unit with the isolation valve closed Use in a CVD gas atmosphere The ambient temperature of the g...

Страница 15: ...is thin at 25 m so avoid use in a location with a large amount of vibration as much as possible The biggest cause of filament breaks is from mechanical shock so use caution regarding the installation...

Страница 16: ...asurements are dependent on the type of gas An environment with a large quantity of fouling Assume internal gauge head contamination Replace the gauge head Pressure does not reach 0 4 Pa or lower even...

Страница 17: ...fer Cause Action The gauge head is dirty Sensitivity has dramatically deviated Replace the gauge head with another gauge head and check the phenomena If there is no problem with the other gauge head t...

Страница 18: ...rature of the gas molecule after the collision If the energy transfer on the filament surface was complete 1 The value of has complicated properties that change according to geometric roughness the pr...

Страница 19: ...5 3 7 7 5 1 4 3 6 7 4 3 5 3 4 7 4 1 2 9 3 7 10 9 3 1 5E 1 9 4 8 5 1 3E 1 8 0 1 1E 1 9 8 7 4 5 6 7 2 20 1 7E 1 2 7E 1 2 0E 1 1 9E 1 2 9E 1 1 7E 1 2 1E 1 2 1E 1 1 5E 1 1 2E 1 1 9E 1 50 3 5E 1 6 0E 1 5...

Страница 20: ...r to the local ULVAC representatives for repair Return charge shall be paid by Buyer Disclaimer 1 Failure occurred after expiration of warranty period 2 Failure caused by force majeure such as fire st...

Страница 21: ...contents of hazardous substances or elements Name of parts Hazardous substances or elements Pb Hg Cd Cr6 PBB PBDE Printed Circuit Board X O O O O O Chassis O O O O O O Connector O O O O O O Label O O...

Страница 22: ...15 9 Dimensions 9 1 SPU dimensions 40 51 113 82 30 R40...

Страница 23: ...16 9 2 Gauge head dimensions 18 NW16 96 102 5...

Страница 24: ...17 9 3 Unit cable SPU P1 RJ 45 P2 RJ 45 WIRE COLOR 1 1 WHITE ORANGE 2 2 ORANGE 3 3 WHITE GREEN 6 6 GREEN 5 5 WHITE BLUE 4 4 BLUE 7 7 WHITE BROWN 8 8 BROWN Shell Shell DRAIN...

Отзывы: