SK00-9663-E8-003-00
The SW100-A maintains the compatibility (OUTPUT) shown in this chapter for the
measured output voltage. By using this function, it can be replaced with SW100-A
without changing the setting to convert the measured output voltage from SW100-A
to the pressure value on the equipment side.
The setting of this function can be easily set with the dedicated application
"UL-MOBI".
OUTPUT
Compatible models/conversion formula
SP1
ULVAC Pirani vacuum gauge sensor unit SP1
PSG
P=10^((V-3.572)/1.286)
(P
:
Pressure[Pa]
、
V
:
Measured output voltage[V])
APG
P=10^(V-4) (P
:
Pressure[Pa]
、
V
:
Measured output voltage[V])
After purchase, please check the instruction manual of SW100 for both new
installation and replacement, and set and connect. When used as a compatible unit,
the performance of SW100-A may be limited, so please read the instruction manual.
※
"UL-MOBI" can be downloaded from our homepage or Google Play. For details,
refer to the SW100 instruction manual.
※
To set SW100, a Windows/Android device with the application "UL-MOBI" installed
and a USB cable (SW100 side: Type-C) are required.
※
UL-MOBI for Windows requires a Windows device with a USB device driver
installed. For details, please refer to the SW100 instruction manual.
※
ULVAC display unit ISG1 does not support the measured value output voltage
conversion formula of PSG and APG.
※
If the specifications of the I/O connector are different, you need to change
the connector.
※
If you set the SW100 on the equipment side, you do not need to use the
compatibility function in this section.