Ultrasound Technologies Ltd
Electromagnetic Compatibility
Fetatrack310 Service Manual Issue3 Page
17
©Ultrasound Technologies Ltd, Lodge Way, Portskewett, Caldicot, South Wales NP26 5PS, United Kingdom.
T +44 (0) 1291 425425 F: +44 (0) 1291 427093 E: [email protected] W:
www.doppler.co.uk
Conductive
disturbance induced
by applied RF field
Test applied a 3Vrms RF magnetic
field to transducer cables with a 2Hz
modulation.
Effect was a displayed rate of 115 to
125 bpm at each harmonic and sub
harmonic of the transducer frequency.
No disturbance was detected at other
frequencies
Applied test signal is very high for high
sensitivity electronics and non applied
transducers. With correctly applied
transducers interference from in band
RF signals is unlikely.
Radiated RF
Test applied: 3V/m 80Mz to 2.5GHz
Effect was a disturbance to the UA
transducer causing a static UA
reading of up to 9 units
Normal operation is unaffected and the
static reading can be cancelled by
pressing the toco zero button
Electrical fast
transients and bursts
Test applied: +/-2KV AC power, +/-
1KV Signal Cables
Effect was a FHR reading of 198 BPM
.
Normal mains power is unlikely to cause
such a transient / burst. Displayed rate is
unlikely to occur when transducers are
connected to a patient.
Electro Static
Discharge
Test applied: +/-2KV, +/-4Kv, +/-
8KV Air Discharge, +/-2KV, +/-
4KV, +/-6KV Contact Discharge.
Repetition Rate 1second
Effect was a FHR reading of 58 BPM .
Unit should be used in a low static
environment. Displayed rate is unlikely
to occur when transducers are
connected to a patient.
Surge
Test applied: +/-0.5KV, +/-1KV, +/-
2KV AC power line to ground, +/-
0.5KV, +/-1KV, +/-6KV AC power
line to line
Effect was a FHR reading of 58 BPM .
Normal mains power is unlikely to cause
such a surge. Displayed rate is unlikely
to occur when transducers are
connected to a patient.
Guidelines for Identifying and resolving adverse EMC conditions
Emissions
Care has been taken through the design and manufacturing processes to minimise
the EM emissions that may be produced by this equipment. However, in the unlikely
event that the unit causes an EM disturbance to adjacent equipment, we suggest that
the procedure is carried out 'out of range' of the affected equipment.
Immunity
If the user has any doubt regarding the unit's EM immunity during routine operation,
we suggest that the source of EM disturbance is identified and its emissions reduced.
If the user has any doubt regarding the identification and resolution of adverse EM
conditions, they may contact Ultrasound Technologies Ltd to seek advice
EMC Testing
During conformity testing the Fetatrack 310 was subjected to International Standard EMC
tests. During the majority of these tests no non conformances were observed.
During EN60601-1-1:2001 testing the FetaTrack 310 was shown to be susceptible to the
following tests.
Содержание Fetatrack 310
Страница 2: ......
Страница 43: ......
Страница 49: ......
Страница 51: ......
Страница 52: ......
Страница 56: ...Q12 collector Unit not printing Q12 collector Unit printing U28 pin 1 U28 pin 2 U28 pin 4 U28 pin 5...
Страница 57: ...L12 pin 2 VCC U29 pin 1 U29 pin 2 U29 pin 4 U29 pin 5...
Страница 58: ...L14 pin 2 15V Supply Rail D22 pin 2 D23 pin 1 Junction L15 pin 1...
Страница 59: ...Q14 Collector Q14 Base Q14 Emitter 15V Supply Rail...
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......