Operating the
FETATRACK 310
In this section, information is supplied which will help you use the
FETATRACK 310
for the fist
time.
User Configuration
Firstly connect the
AC supply cord.
It is necessary to set the operation of the
FETATRACK 310
to meet your requirements.
Before turning the
FETATRACK 310
on, press and hold down the
Toco Zero
button. Then, whilst
keeping the
Toco Zero
button pressed down, turn the
FETATRACK 310
on by flicking the
AC
input on/off
switch which is located on the rear of the unit as part of the
AC line input socket
.
When the unit is on the front panel LED
will be illuminated.
The
FETATRACK 310
starts and then enters its
User Configuration
Mode, release
the Toco Zero button as soon as ‘
Calibrate System’
is displayed .
You are then prompted to enter the date if different to that displayed.
To change the day and month press the
Volume Up
button. One press advances the
day by one, pressing the
Volume Up
button for longer advances the day by 10.
To change the year press the
Volume Down
button. One press advances the year by
one, pressing the
Volume Down
button for longer advances the year by 10.
When the correct date is displayed press the
Toco Zero
button once.
You are then prompted to enter the time if different to that displayed.
This works in a similar manner to the date with the
Volume Up
button advancing the
hours and the
Volume Down
button advancing the minutes.
When the correct time is displayed press the
Toco Zero
button once.
You are then prompted to enter the Toco Base Line offset.
This sets an artificial zero line for the toco transducer above zero, it can be set by
pressing the
Volume Up or Volume Down
buttons for a value between 0 - 20.
After setting this value the Toco will be set to this every time the
Toco Zero
switch is
pressed, and allows small negative Toco excursions to be seen on the chart.
When the chosen zero offset is displayed press the
Toco Zero
button once.
Содержание Fetatrack 310
Страница 2: ......
Страница 43: ......
Страница 49: ......
Страница 51: ......
Страница 52: ......
Страница 56: ...Q12 collector Unit not printing Q12 collector Unit printing U28 pin 1 U28 pin 2 U28 pin 4 U28 pin 5...
Страница 57: ...L12 pin 2 VCC U29 pin 1 U29 pin 2 U29 pin 4 U29 pin 5...
Страница 58: ...L14 pin 2 15V Supply Rail D22 pin 2 D23 pin 1 Junction L15 pin 1...
Страница 59: ...Q14 Collector Q14 Base Q14 Emitter 15V Supply Rail...
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......