UltraSonic LLC. Model Ultra 2400FA
9
Button until Time Setting use Up/Down Buttons to set Year
press Menu Button to save press Up/Down Buttons to set
Month, press Menu Button to save press Up/Down Buttons
to set Day, press Menu Button to save press Up/Down
Buttons to set Hour, press Menu Button to save press
Up/Down Buttons to set Minute, press Menu Button to save
press Up/Down Buttons to set Seconds, press Menu
Button to save press Up/Down Buttons to set Day, press Menu
Button to Save. Press UP Button to move to next setting.
•
Exit:
Press menu key to Exit
NORMAL OPERATION
Make sure solution is at operating temperature.
Arrange parts in baskets & lower basket into solution.
Push green
Cycle Start
button on the front of the machine.
Close lid when not loading/unloading parts.
When ultrasound turns OFF,
Filtration will turn on
automatically to skim off any
surface oils & contaminants into the weir tank.
Remove basket.
NOTE:
Parts are hot, use proper protective equipment to prevent
exposure.
Do Not Plug in Machine until unit is full of water and cleaning solution, doing so
may cause damage to your UltraSonic tank and Void warranty
MAINTENANCE
Preventative - There is a minimum of preventive maintenance required. The overall
environment around the unit needs to be kept clean, dry and dust free.
Routine -
Filter Element
should be replaced when the pressure gauge indicates rise in
pressure or the flow from the spray bar located in the front of the main tank is slowed.
Most users regularly replace their filter each month to avoid any decrease of flow.
Solution should be replaced regularly to avoid a drop in cleaning performance that
will result from too many contaminants in the solution.
NOTE:
Canister and solution
are hot and may be contaminated; protective clothing and equipment should be used
to prevent any exposure. To change filter elements, remove solution from
front/cleaning tank until water level is below spray bar. Remove cover in back of
machine to access filter canister using a 15/16" wrench to loosen and remove canister.