Turn aluminum handpiece
shell counter-clockwise
to be removed.
Remove aluminum handpiece
shell for cleaning and sterilization.
7mm PBM Tip turn counter clockwise to remove it.
3mm PBM Tip turn counter clockwise to remove it.
Use extreme care not to accidentally
damage LED lens.
CLEANING AND STERILIZATION PROCEDURES
NOTE:
The exterior of the laser unit is not routinely contaminated by procedures. The Guided Touch Interfce and
Electroluminescent display should be covered with a protective clear adhesive barrier film, replaceable after each patient.
If the exterior of the laser unit should become contaminated, it should be wiped down with CaviWipes, or equivalent product,
then re-covered with a new protective plastic cover. We recommend wringing out any cleaning wipes before use to avoid
dripping liquid on the laser unit.
To reassemble, slide the handpiece shell onto the handpiece body and turn clockwise to tighten.
Remove / reassemble the handpiece shell or 3mm/7mm PBM adapter following the instructions below.
DO NOT
SPRAY ANY DISINFECTANT DIRECTLY ON THE LASER UNIT, BECAUSE IT WILL DAMAGE
THE TRANSPARENT ELECTROLUMINESCENT DISPLAY.
DO NOT
USE ABRASIVE MATERIALS TO CLEAN THE LASER OR THE DISPLAY.