Page
14
of
20
Issue 9
– 17122020
12.3 Power unit disinfection
The power unit can be cleaned by wiping down with a cloth dampened with hot water at 60
0
C
containing detergent or with sodium hypochlorite (up to 10,000 parts per million available chlorine).
12.4 Cover laundering
The cushion cover and air pipe cover can also be machine washed. Cushion covers and air pipe
cover should be completely removed prior to laundering.
Where required cushion covers can be laundered in a pre-wash at 60
0
C for up to 15 minutes and in a
main wash at 84
0
C for up to 15 minutes. This should be followed by a cold rinse and extraction.
However it is recommended that you check your local policy to determine the time/temperature ratio
required to achieve thermal disinfection.
Cushion covers may be tumble dried or air dried. They may be tumble dried on a low heat for up to 90
minutes. Drying temperature must not exceed 40
0
C.
The cushion cover and cushion must be dry prior to refitting.
Caution: Exceeding the temperature can cause significant damage to the cushion
cover.
13 Storage
The cushion replacement should be loosely rolled lengthwise with the cover innermost, taking care
not to strain the air pipe. It should then be placed in in a suitable protective cover with the power unit
and stored in an area appropriate for electronic medical devices.
Disconnect the air hose connector to release the air.
The power cord could be wrapped around the power unit bumper or disconnected for storage.
Caution: Do not fold or crease cushions