Table 2: Alert/Error Code Reference Table
Priority
High
↓
Low
Warning
Code
Indicator
LED
Audible Output
Mode
Condition of
Output
Warning Description
Remarks
0
N/A
N/A
ONCE
Not in System
Shutdown
Key Tone
Key Tone from
Functional Button
1
Power Failure
ONCE
Power-Off
System Shutdown
Shutdown
2
ALL LED
ONCE
Operate or
Standby
Power-On
All Indicators
Light On
3
N/A
N/A
ONCE
Operate or
Standby
State / Mode
Switching
No Display
4
Auto-Firm
ONCE
Operate
Mattress Inflation
Completion
Inflation Ended
5
Auto-Firm
ONCE
Operate
Auto-Firm Completion
Auto-Firm Ended
6
Static
ONCE
Operate
Static Completion
Static Ended
7
N/A
Power Failure
REPEAT
(Cycle 4 sec.)
Power-Off
Power Failure Alarm
No Display
8
Low Pressure
REPEAT
(Cycle 4 sec)
Operate or
Standby
Power-On Inflation Failure
Alarm
Inflate Failure
9
Low Pressure
REPEAT
(Cycle 4 sec)
Operate or
Standby
Auto-Firm
Failure Alarm
Auto-Firm Failure
10
Low Pressure
REPEAT
(Cycle 4 sec)
Operate or
Standby
Low Pressure
Overtime Alarm
Low Pressure
11
Service
REPEAT
(Cycle 4.5 sec)
Operate or
Standby
High Pressure
Overtime Alarm
High Pressure
12
Service
REPEAT
(Cycle 4.5 sec)
Operate or
Standby
High Ambient
Temperature Alarm
High Temperature
13
Service
REPEAT
(Cycle 4.5 sec)
Operate or
Standby
Air Valve 1 Positioning
Failure Alarm
Air Valve 1 failure
14
Service
REPEAT
(Cycle 4.5 sec)
Operate or
Standby
Air Valve 2 Positioning
Failure Alarm
Air Valve 2 failure
15
Service
REPEAT
(Cycle 15 sec)
Operate or
Standby
Battery Low Alarm
Battery would need
to be replaced
16
NONE
NONE
Factory
Calibration
Mode
Calibration Not Completed
Calibration
Unfinished
17
NONE
NONE
Factory
Calibration
Mode
Calibration
Completed
Calibration
Completed