background image

 

Page 

14

 of 

28

 

Issue 5: 07/01/2021 
 
 

9.5  Function mode switch 

 
 

9.5.1  Alternate 

 

 

Alternating

 mode is the default mode for the system. Within this mode the mattress will 

operate in an alternating 1-in-3 cell cycle. The alternating cycle will continue at the 

selected cycle time until another mode is selected. 

 

 

9.5.2  Static 

 

Pressing the 

Function Mode Selection

 button until the 

Static 

indicator illuminates puts 

the system into 

Static

 mode.  

 

Within this mode the mattress will maintain selected constant pressure. After 20 minutes 

the system will automatically revert back to 

Alternating

 mode.  

 
 

9.5.3  Constant Low Pressure (CLP) 

 

 

Pressing the 

Function Mode Selection

 button until the 

CLP 

indicator illuminates puts 

the system into 

Constant Low Pressure (CLP)

 mode.  

 

Within this mode the internal cell pressures within the mattress will be reduced to 

15mmHg ± 1mmHg, this allows the patient to sink further into the mattress, the mattress 

will provide a comfortable static surface. 

 

 

9.6  Auto-Firm 

 

 

Auto-Firm 

mode can be selected by pressing the 

Auto-Firm 

button. In this mode all 

cells inflate to a single pressure setting to provide a firm and stable surface for nursing 

procedures or for patient ingress/egress. The system will automatically return to the 

Last 

Mode of Operation

 at the previously selected comfort level after 20 minutes. 

 

9.7  Alternate cycle time selection  

 
 

 

The alternating cycle times can be selected to provide an individualised care program for 

each patient.  

 

Within 

Alternating

 mode the alternating cycle time can be selected by pressing the 

Cycle

 button. Selections can be made from 10-30 minutes at 5 minute intervals. The 

cycle time will be displayed in the LED screen. 

 

 

 

 

Содержание Tamora III Advance

Страница 1: ...Tamora III Advance Dynamic Mattress Replacement System UPRA367803A 3 User Manual Issue 5 07 01 2021 UPRA367803A 3 UM 5...

Страница 2: ...Page 2 of 28 Issue 5 07 01 2021...

Страница 3: ...standby 12 9 2 Comfort level 13 9 3 Auto detection 13 9 4 Control panel lockout 13 9 5 Function mode switch 14 9 5 1 Alternate 14 9 5 2 Static 14 9 5 3 Constant Low Pressure CLP 14 9 6 Auto Firm 14 9...

Страница 4: ...Page 4 of 28 Issue 5 07 01 2021 16 1 General 21 16 2 Fuse replacement 21 16 3 Air filter replacement 21 17 EMC Information 22...

Страница 5: ...is for the prevention and treatment of pressure ulcers while optimising patient comfort for patients up to 255kg 40 stone 2 1 Contraindications The mattress is not suitable for use on patients with u...

Страница 6: ...01 2021 Mattress Replacement CPR Valve Cable Management Fixings Air Outlet port 4 x Static Head Cells TPU 20 x Alternating Cells TPU Removable Cover Figure of 8 cell technology FOOT END HEAD END Knitt...

Страница 7: ...in order to prevent personal injury BF symbol indicates this product is according to degree of protection against electric shock for type BF equipment The operator must read this document User Manual...

Страница 8: ...of the power cable e g by kinking shearing or other mechanical damages The power cable for this product must be unplugged from the mains power outlet socket and suitably stowed before moving cleaning...

Страница 9: ...cation Class II Type BF Warranty 2 year Operation environment 50C to 400C 15 RH 93 RH no condensation Storage environment 25 70 93 RH no condensation Environment pressure 70 kPa 101 3 kPa Water dust p...

Страница 10: ...ewed from the foot of the bed There are security straps fitted to the base cover of the mattress which should be fastened loosely to convenient points on the bedframe patient surface Caution Ensure wh...

Страница 11: ...connector to the Pump and then ensure that the air tubes are free from any kind of obstruction and are not kinked Plug the power cable into a suitable electrical socket and switch ON using mains power...

Страница 12: ...m Indicator 2 Comfort Setting Indicators 6 Operate Standby Button Indicators 3 Control Panel Lockout Button Indicator 7 Cycle Time Selection Button Alternate 4 Function Mode Selection Button Alternate...

Страница 13: ...Once activated the indicator will flash and the system will start detecting and will set the internal pressure for the appropriately The indicator light will flash until the patient s weight and posi...

Страница 14: ...ssure CLP mode Within this mode the internal cell pressures within the mattress will be reduced to 15mmHg 1mmHg this allows the patient to sink further into the mattress the mattress will provide a co...

Страница 15: ...p the air release The mattress deflates in 10 15 seconds If re inflating the mattress ensure the dial of the CPR attachment is rotated until it clicks into the CLOSED position 11 Transport Mode If the...

Страница 16: ...hout turning the Pump OFF an audible alarm will be heard and the Power Failure indicator will illuminate The audible alarm may be cancelled by pressing the Alarm Mute button Note When the Pump has not...

Страница 17: ...e 4 sec Operate or Standby Power On Inflation Failure Alarm Inflate Failure 9 Low Pressure REPEAT Cycle 4 sec Operate or Standby Auto Firm Failure Alarm Auto Firm Failure 10 Low Pressure REPEAT Cycle...

Страница 18: ...h is turned ON The Pump is operating but the mattress is not alternating Ensure that the mattress inflation process is complete Check that the Alternate indicator on the control panel is illuminated I...

Страница 19: ...water Caution When cleaning disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents or dis...

Страница 20: ...ied or air dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceeding th...

Страница 21: ...able and plug if there are abrasions or excessive wears 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are stubbed together correctly 3 Check the air hoses for any...

Страница 22: ...d professional healthcare environment RF emissions CISPR 11 Group 1 The device s uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any i...

Страница 23: ...nvironment Surge IEC 61000 4 5 0 5kV 1kV line s to line s 0 5kV 1kV 2kV line s to earth 0 5kV 1kV line s to line s Not applicable Mains power quality should be that of a typical home and professional...

Страница 24: ...ands between 0 15 MHz and 80 MHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Portable and mobile RF communications equipment should be used no closer to any part of the device s including cabl...

Страница 25: ...t power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0...

Страница 26: ...0 2 0 3 9 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18 Hz 2 0 3 28 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMT...

Страница 27: ...Page 27 of 28 Issue 5 07 01 2021 Notes...

Страница 28: ...ue 5 07 01 2021 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimate...

Отзывы: