(2) Check mattress cover for signs of wear or damage. Ensure mattress cover and tubes are
stubbed together correctly.
(3) Check the air hoses for any kink or break. For replacement, please contact Ultimate
Healthcare or your local distributor.
13.2 Fuse replacement
(1) Disconnect the plug from mains power when a blown fuse is suspected.
(2) Remove the cover of the fuse holder by means of a small screwdriver.
(3) Insert a new fuse of the correct rating in, and replace the cover of the fuse holder back. The
fuse should be rated as T1a/250V.
13.3 Air filter replacement
(1) Replace the air filter located at the handle on the back of the power unit.
(2) The filter is reusable and can be washed gently with a mild detergent and water. Dry the filter
before use.
(3) Check and replace air filter regularly if environment is dirty.
14 . EMC Guidance and Declaration
Alternating Pressure Mattress Pump, Model No: HF609
EMC Compliant with and tested to:
EN55011:2009
EN60601-1-2:2015
EN61000-3-2:2014
EN61000-3-3:2013
Test Report & reference No:
LCS171128004CE
LCSEMC-1.0
Electrical Ratings:
220-240V AC
50Hz
70mA
7W