8.5  Alarm mute  

 

 

The 

Alarm Mute

 button temporarily resets the audible 

Low Pressure/Power 

Failure/Service

 alarms. Should the situation not be resolved and the fault condition 

continues the alarm will resume notifying carer. 

9. Alarms & Fault Findings 

 

The system is equipped with audible and visual alarm indicators. These alert the user to the status of 

the available mains supply and any mattress defect. 
 

9.1  Low pressure alarm 

 

Upon detection of low pressure, an audible alarm will be heard and the 

Low Pressure

 indicator will 

illuminate. The audible alarm may be cancelled by touching the 

Alarm Mute

 button.

 

The 

Low 

Pressure

 indicator will continuously illuminate until the low pressure fault condition is resolved.  

 

This condition could be caused, for example, by incorrect fitting of the air pipe connector, opening of 

the CPR valve or a leak in the mattress due to a cut or puncture. 
 

9.2  Power failure alarm 

 

If at any time the mains power should be removed from the power unit or the power cable is 

unplugged without turning the power unit OFF, an audible alarm will be heard and the 

Power Failure

 

indicator will illuminate. The audible alarm may be cancelled by touching the 

Alarm Mute

 button. 

 

The mattress is capable of supporting the patient in static mode for a period of time. 
 
 
 

 

 

 

 

 

Содержание Diversi-Foam II

Страница 1: ...Diversi Foam II Hybrid Mattress Replacement System UPRA3678H 4 and UPRA4278H 4 ________________________________________________ User Manual Issue 1 03 10 2019 UPRA3678H 4 UPRA4278H 4 UM 1...

Страница 2: ...8 2 Comfort level 10 8 3 Control panel lockout 10 8 4 Function mode switch 10 8 4 1 Alternate 10 8 4 2 Constant Low Pressure CLP 10 8 5 Alarm mute 11 9 Alarms Fault Findings 11 9 1 Low pressure alarm...

Страница 3: ...time The information in this User Manual may vary slightly with respect to the basic design of the product 2 Intended Use The intended use of this product is to prevent and or manage pressure ulcers w...

Страница 4: ...Page 4 of 16 Issue 1 03 10 2019 Mattress Replacement Power Unit Welded mattress cover U core foam Foot cells Static head Body cells...

Страница 5: ...icates potential danger that requires correct procedures or practises in order to prevent personal injury The operator must read this document User Manual before use IP21 Water and dust protection cla...

Страница 6: ...caused by inappropriate handling of the power cable e g by kinking shearing or other mechanical damages The power cable for this product must be unplugged from the mains power outlet socket and suita...

Страница 7: ...rranty 2 years Operation environment 50C to 400C 15 RH 93 RH no condensation Storage environment 25 70 93 RH no condensation Environment pressure 70 kPa 101 3 kPa Water dust protection classification...

Страница 8: ...nnot be twisted trapped crushed or stressed 7 3 Connecting mattress to the power unit Remove the cover of the air pipe connector and connect the air pipe connector to the power unit and then ensure th...

Страница 9: ...ustments to the mattress and also indicates fault conditions service requirements These are either visual indicator lights or audible 1 Power Button 5 CLP Button 2 Pressure Up Button 6 Weight Pressure...

Страница 10: ...er mattress 8 3 Control panel lockout If the control panel is not used for a period of 15 seconds it will lock out and inhibit the use of the functions Additionally if you wish to lock out the control...

Страница 11: ...l illuminate The audible alarm may be cancelled by touching the Alarm Mute button The Low Pressure indicator will continuously illuminate until the low pressure fault condition is resolved This condit...

Страница 12: ...of above steps have been checked Press Alarm Mute for system to be verified again Power Failure Alarm Indicator is flashing and sounding Check the power unit is connected to the mains power supply and...

Страница 13: ...r Caution When cleaning disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents or disinfec...

Страница 14: ...dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceeding the temperatu...

Страница 15: ...e correct rating in and replace the cover of the fuse holder back The fuse should be rated as T1a 250V 13 3 Air filter replacement 1 Replace the air filter located at the handle on the back of the pow...

Страница 16: ...ue 1 03 10 2019 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimate...

Отзывы: